Suppr超能文献

体外冲击波碎石术对患者的辐射剂量。

Radiation doses to patients from extracorporeal shock wave lithotripsy.

作者信息

Sandilos Panagiotis, Tsalafoutas Ioannis, Koutsokalis Georgios, Karaiskos Pantelis, Georgiou Evangelos, Yakoumakis Emanouil, Vlahos Lampros

机构信息

Department of Radiology, Areteion Hospital, Medical School, University of Athens, 76 Vasilissis Sophias, 115 28 Athens, Greece.

出版信息

Health Phys. 2006 Jun;90(6):583-7. doi: 10.1097/01.HP.0000196112.94905.dd.

Abstract

The aim of this study was to determine the radiation doses to patients during extracorporeal shock wave lithotripsy (ESWL) and compare them with the available bibliographical data. In this method localization of the renal stones is attained by the use of fluoroscopy, and thus ESWL is included among those medical practices associated with patient radiation exposure. The entrance surface dose was measured using 3-4 thermoluminescent dosimeters positioned on the patients' backs at the entrance surfaces of the two x-ray beams for 50 ESWL procedures. Fluoroscopy time and number of spot films were also recorded. The average entrance surface dose at the patient's side with the renal stone was estimated to be 76.5 mGy for the oblique x-ray beam and 44.5 mGy for the PA x-ray beam. The mean fluoroscopy time was 204 s while 4 spot films were acquired on average. The mean effective dose (E) was estimated as 1.63 mSv per patient. The mean entrance surface dose values recorded in this study are comparable to but smaller than the values reported in the literature for ESWL, while the mean fluoroscopy time is within the range of values reported by other authors. On the other hand, the estimated E value is relatively higher compared to the corresponding values given in the literature based on patient measurements. However, it is comparable to recent published data acquired using extended measurements in an anthropomorphic phantom.

摘要

本研究的目的是确定体外冲击波碎石术(ESWL)期间患者所接受的辐射剂量,并将其与现有的文献数据进行比较。在这种方法中,肾结石的定位是通过荧光透视实现的,因此ESWL属于与患者辐射暴露相关的医疗操作。使用3 - 4个热释光剂量计,放置在患者背部两个X射线束的入射表面,对50例ESWL手术进行入射表面剂量测量。同时记录荧光透视时间和点片数量。对于斜位X射线束,患侧患者的平均入射表面剂量估计为76.5 mGy,对于正位(PA)X射线束为44.5 mGy。平均荧光透视时间为204秒,平均采集4张点片。估计每位患者的平均有效剂量(E)为1.63 mSv。本研究记录的平均入射表面剂量值与文献中报道的ESWL值相当,但低于这些值,而平均荧光透视时间在其他作者报道的值范围内。另一方面,与基于患者测量的文献中给出的相应值相比,估计的E值相对较高。然而,它与最近使用人体模型扩展测量获得的已发表数据相当。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验