Sakai Rina, Itoman Moritoshi, Mabuchi Kiyoshi
Department of Biomedical Engineering, School of Allied Health Science, Kitasato University, 1-15-1 Kitasato, Sagamihara City, Kanagawa 228-8555, Japan.
Clin Biomech (Bristol). 2006 Oct;21(8):826-33. doi: 10.1016/j.clinbiomech.2006.03.008. Epub 2006 May 15.
It is now recognized that initial stability is essential for avoidance of thigh pain in hip replacement. The initial stability corresponds to an optimal stress distribution of cementless orthopedic implants. Although the relationship between the contour and stress at the fixation site has been analyzed, guidelines on stem design have not been established.
Finite element models of three currently-used stems were constructed for a computer simulation. Contact stress at the fixation site of a joint prosthesis was analyzed by an explicit three-dimensional finite element method. The stress immediately after applying load using a film or sensor which can measure contact stress was observed. The situation of the initial fixation about the specific part which becomes important clinically based on the results was clarified.
We introduced fluctuation area as a measure to evaluate the primary fixation of femoral stems. It was found that the stress distribution on the PerFix SV stem fluctuated with a slight disturbance. On the Intra-Medullary Cruciate stem, the high stress areas were distributed on the proximal area and under the pin. The high stress area on the VerSys stem were spread on the medial side.
This study highlights the mechanical instability of the fixation site of joint prostheses, and thus suggests that the general idea that unconditionally recommends a larger area for the fixation site of joint prostheses should be revised.
目前人们认识到,初始稳定性对于避免髋关节置换术后大腿疼痛至关重要。初始稳定性对应于非骨水泥型骨科植入物的最佳应力分布。尽管已经分析了固定部位的轮廓与应力之间的关系,但尚未制定关于柄部设计的指南。
构建了三种当前使用的柄部的有限元模型用于计算机模拟。采用显式三维有限元方法分析关节假体固定部位的接触应力。使用可测量接触应力的薄膜或传感器观察加载后立即产生的应力。根据结果阐明临床上重要的特定部位的初始固定情况。
我们引入波动面积作为评估股骨干初始固定的指标。发现PerFix SV柄部的应力分布在轻微干扰下波动。在髓内十字柄上,高应力区域分布在近端区域和销钉下方。VerSys柄部的高应力区域分布在内侧。
本研究突出了关节假体固定部位的机械不稳定性,因此表明应修订无条件推荐关节假体固定部位采用更大面积的普遍观念。