Bedogni R, Angelone M, Esposito A, Chiti M
ENEA-ION-Radiation Protection Institute, 16 via Dei Colli, I-40136 Bologna, Italy.
Radiat Prot Dosimetry. 2006;120(1-4):369-72. doi: 10.1093/rpd/nci577. Epub 2006 May 18.
In the framework of collaboration among the ENEA Radiation Protection Institute (Bologna), the ENEA Fusion Department (Frascati) and the INFN-LNF-Radiation Protection Group (Frascati), an experimental campaign was organised on the usage of thermoluminescence dosemeters (TLDs) for the dosimetric and spectrometric characterisation of neutron fields. Commercially available TLDs of different material and different sensitivity to photons and thermal neutrons were selected, namely TLD600H and TLD700H from Harshaw, GR206 and GR207 from SSDML (China), MCP-6s from TLD Poland. The detectors were first calibrated in standard fields of photons ((60)Co) and thermal neutrons at the ENEA-IRP Secondary Standard Calibration Laboratory of Bologna, then exposed in fast neutron standard fields of different energy, using a standard multisphere moderating assembly. The paper compares the dosimetric characteristics of the studied TL detectors, underlining the (n-gamma) discrimination capability, and discusses their spectrometric performances addressed to radiation protection applications.
在意大利国家新技术、能源与可持续经济发展局辐射防护研究所(博洛尼亚)、意大利国家新技术、能源与可持续经济发展局聚变部(弗拉斯卡蒂)和意大利国家核物理研究所-弗拉斯卡蒂国家实验室-辐射防护组(弗拉斯卡蒂)的合作框架下,组织了一场关于使用热释光剂量计(TLD)对中子场进行剂量学和能谱学表征的实验活动。选择了不同材料且对光子和热中子具有不同灵敏度的市售TLD,即哈肖公司的TLD600H和TLD700H、中国固态剂量监测有限公司的GR206和GR207、波兰TLD公司的MCP-6s。这些探测器首先在博洛尼亚的意大利国家新技术、能源与可持续经济发展局辐射防护二级标准校准实验室的光子(钴-60)和热中子标准场中进行校准,然后使用标准多球慢化组件在不同能量的快中子标准场中进行照射。本文比较了所研究的热释光探测器的剂量学特性,强调了(n-γ)鉴别能力,并讨论了其针对辐射防护应用的能谱性能。