Chevalier Margarita, Morán Pilar, Ten José I, Fernández Soto José M, Cepeda T, Vañó Eliseo
Department of Radiology, Facultad de Medicina, Universidad Complutense de Madrid, 28040 Madrid, Spain.
Med Phys. 2004 Sep;31(9):2471-9. doi: 10.1118/1.1784591.
In the present investigation, we analyze the dose of 5034 patients (20,137 images) who underwent mammographic examinations with a full-field digital mammography system. Also, we evaluate the system calibration by analyzing the exposure factors as a function of breast thickness. The information relevant to this study has been extracted from the image DICOM header and stored in a database during a 3-year period (March 2001-October 2003). Patient data included age, breast thickness, kVp, mAs, target/filter combination, and nominal dose values. Entrance surface air kerma (ESAK) without backscatter was calculated from the tube output as measured for each voltage used under clinical conditions and from the tube loading (mAs) included in the DICOM header. Mean values for the patient age and compressed breast thickness were 56 years (SD: 11) and 52 mm (SD: 13), respectively. The majority of the images was acquired using the STD (for standard) automatic mode (98%). The most frequent target/filter combination automatically selected for breast smaller than 35 mm was Mo/Mo (75%); for intermediate thicknesses between 35 and 65 mm, the combinations were Mo/Rh (54%) and Rh/Rh (38.5%); Rh/Rh was the combination selected for 91% of the cases for breasts thicker than 65 mm. A wide kVp range was observed for each target/filter combination. The most frequent values were 28 kVp for Mo/Mo, 29 kVp for Mo/Rh, and 29 and 30 kV for Rh/Rh. Exposure times ranged from 0.2 to 4.2 s with a mean value of 1.1 s. Average glandular doses (AGD) per exposure were calculated by multiplying the ESAK values by the conversion factors tabulated by Dance for women in the age groups 50 to 64 and 40 to 49. This approach is based on the dependence of breast glandularity on breast thickness and age. The total mean average glandular dose (AGD(T)) was calculated by summing the values associated with the pre-exposure and with the main exposure. Mean AGD(T) per exposure was 1.88 mGy (CI 0.01) and the mean AGD(T) per examination was 3.8 mGy, with 4 images per examination on average. The mean dose for cranio-caudal view (CC) images was 1.8 mGy, which is lower than that for medio-lateral oblique (MLO) view because the thickness for CC images was on average 10% lower than that for MLO images. Mean AGD(T) for the oldest group of women (1.90) was 3% higher than the AGD(T) for the younger group (1.85) due to the larger compressed breast thickness of women in the older group (10% on average). Differences between the corresponding AGD(T) values of each age group were lowest for breast thicknesses in the range 40-60 mm, being slightly higher for the women in the older group.
在本研究中,我们分析了5034例患者(20137幅图像)的剂量,这些患者均使用全视野数字乳腺摄影系统进行了乳腺摄影检查。此外,我们通过分析作为乳房厚度函数的曝光因素来评估系统校准。与本研究相关的信息已从图像DICOM头文件中提取,并在3年期间(2001年3月至2003年10月)存储在数据库中。患者数据包括年龄、乳房厚度、千伏峰值(kVp)、毫安秒(mAs)、靶/滤过组合以及标称剂量值。根据临床条件下每个使用电压的管输出以及DICOM头文件中包含的管负载(mAs),计算无反向散射的入射表面空气比释动能(ESAK)。患者年龄和压缩乳房厚度的平均值分别为56岁(标准差:11)和52毫米(标准差:13)。大多数图像是使用STD(标准)自动模式采集的(98%)。对于小于35毫米的乳房,自动选择的最常见靶/滤过组合是钼/钼(Mo/Mo,75%);对于35至65毫米的中等厚度乳房,组合是钼/铑(Mo/Rh,54%)和铑/铑(Rh/Rh,38.5%);对于厚度超过65毫米的乳房,91%的病例选择的组合是铑/铑。观察到每个靶/滤过组合的千伏峰值范围很广。最常见的值是钼/钼为28 kVp,钼/铑为29 kVp,铑/铑为29 kV和30 kV。曝光时间范围为0.2至4.2秒,平均值为1.1秒。每次曝光的平均腺体剂量(AGD)通过将ESAK值乘以Dance为50至64岁和40至49岁年龄组女性列出的转换因子来计算。这种方法基于乳房腺体组织与乳房厚度和年龄的相关性。通过将与预曝光和主曝光相关的值相加来计算总平均腺体剂量(AGD(T))。每次曝光的平均AGD(T)为1.88毫戈瑞(置信区间0.01),每次检查的平均AGD(T)为3.8毫戈瑞,平均每次检查有4幅图像。头尾位(CC)视图图像的平均剂量为1.8毫戈瑞,低于内外斜位(MLO)视图,因为CC视图图像的厚度平均比MLO视图低10%。由于老年组女性的压缩乳房厚度更大(平均10%),老年组女性(1.90)的平均AGD(T)比年轻组(1.85)高3%。每个年龄组相应AGD(T)值之间的差异在乳房厚度40至60毫米范围内最低,老年组女性的差异略高。