Suppr超能文献

孟加拉国采用氢化物发生-原子吸收分光光度法对饮用水中砷含量进行的改进常规分析。

A modified routine analysis of arsenic content in drinking-water in Bangladesh by hydride generation-atomic absorption spectrophotometry.

作者信息

Wahed M A, Chowdhury Dulaly, Nermell Barbro, Khan Shafiqul Islam, Ilias Mohammad, Rahman Mahfuzar, Persson Lars Ake, Vahter Marie

机构信息

Nutritional Biochemistry Department, Laboratory Sciences Division, ICDDR,B: Centre for Health and Population Research GPO Box 128, Dhaka 1000, Bangladesh.

出版信息

J Health Popul Nutr. 2006 Mar;24(1):36-41.

Abstract

The high prevalence of elevated levels of arsenic in drinking-water in many countries, including Bangladesh, has necessitated the development of reliable and rapid methods for the determination of a wide range of arsenic concentrations in water. A simple hydride generation-atomic absorption spectrometry (HG-AAS) method for the determination of arsenic in the range of microg/L to mg/L concentrations in water is reported here. The method showed linearity over concentrations ranging from 1 to 30 microg/L, but requires dilution of samples with higher concentrations. The detection limit ranged from 0.3 to 0.5 microg/L. Evaluation of the method, using internal quality-control (QC) samples (pooled water samples) and spiked internal QC samples throughout the study, and Standard Reference Material in certain lots, showed good accuracy and precision. Analysis of duplicate water samples at another laboratory also showed good agreement. In total, 13,286 tubewell water samples from Matlab, a rural area in Bangladesh, were analyzed. Thirty-seven percent of the water samples had concentrations below 50 microg/L, 29% below the WHO guideline value of 10 microg/L, and 17% below 1 microg/L. The HG-AAS was found to be a precise, sensitive, and reasonably fast and simple method for analysis of arsenic concentrations in water samples.

摘要

包括孟加拉国在内的许多国家,饮用水中砷含量升高的情况普遍存在,因此有必要开发可靠且快速的方法来测定水中多种砷浓度。本文报道了一种简单的氢化物发生-原子吸收光谱法(HG-AAS),用于测定水中微克/升至毫克/升浓度范围内的砷。该方法在1至30微克/升的浓度范围内呈线性,但对于较高浓度的样品需要进行稀释。检测限为0.3至0.5微克/升。在整个研究过程中,使用内部质量控制(QC)样品(混合水样)和加标内部QC样品以及某些批次的标准参考物质对该方法进行评估,结果显示出良好的准确度和精密度。在另一个实验室对重复水样进行分析也显示出良好的一致性。总共对来自孟加拉国农村地区Matlab的13286份管井水样品进行了分析。37%的水样浓度低于50微克/升,29%低于世界卫生组织10微克/升的指导值,17%低于1微克/升。结果发现,HG-AAS是一种精确、灵敏、相当快速且简单的分析水样中砷浓度的方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验