Aljehani Abdulaziz, Bamzahim Mohammad, Yousif Mirgani Awad, Shi Xie-Qi
Department of Cariology and Endodontology, Institute of Odontology, Karolinska Institutet, Huddinge, Sweden.
Oral Health Prev Dent. 2006;4(2):145-50.
The aim of this study was to evaluate the reliability of a laser-induced infrared fluorescence method, DIAGNOdent, for measuring orthodontically induced white spot lesions. The subjects comprised 13 orthodontic patients, aged 13-17 years, who had recently completed fixed appliance therapy: 137 test teeth were selected, with white spot lesions on the facial or buccal smooth surfaces. An initial visual inspection was performed to localise and record the measuring region. The predetermined measuring regions were scanned to locate the sites of the highest reading. The readings and their corresponding sites were registered on the print out photographs. Following the measurement by the first examiner, the second and the third examiners took DIAGNOdent readings independently at the same lesion sites indicated on the photographs, under identical conditions. One week later, DIAGNOdent readings of the same lesions were retaken by the three observers working independently. Intra- and inter- examiner agreements on DIAGNOdent quantification of lesion severity were analysed by Intra-class correlation coefficient (ICC). The ICC values for intra-examiner agreement for the three examiners were 0.91, 0.97, and 0.98, respectively, with a mean value of 0.95, indicating excellent agreement. The ICC values for inter-examiner agreement were comparatively lower: 0.69 and 0.82 for the first and second measurements, respectively. It was concluded that the reliability of the DIAGNOdent readings on white spot lesions associated with orthodontic banding was good.
本研究的目的是评估激光诱导红外荧光法(DIAGNOdent)测量正畸性白斑损害的可靠性。研究对象包括13名年龄在13至17岁之间、近期完成固定矫治器治疗的正畸患者:共选取137颗测试牙,其唇面或颊面光滑表面有白斑损害。首先进行视觉检查以定位并记录测量区域。对预先确定的测量区域进行扫描,以找出读数最高的部位。将读数及其相应部位记录在打印出的照片上。在第一位检查者测量之后,第二位和第三位检查者在相同条件下,独立对照片上指示的相同损害部位进行DIAGNOdent读数。一周后,由三位观察者独立重新测量相同损害的DIAGNOdent读数。采用组内相关系数(ICC)分析检查者内和检查者间对DIAGNOdent量化损害严重程度的一致性。三位检查者的检查者内一致性ICC值分别为0.91、0.97和0.98,平均值为0.95,表明一致性极佳。检查者间一致性的ICC值相对较低:第一次和第二次测量分别为0.69和0.82。得出的结论是,DIAGNOdent对与正畸带环相关的白斑损害读数的可靠性良好。