J Pharm Biomed Anal. 1984;2(3-4):519-26. doi: 10.1016/0731-7085(84)80055-2.
Felodipine in plasma was extracted with toluene and determined by automated capillary gas chromatography with electron capture detection. Undesirable oxidation of the dihydropyridine derivative in the injector and in the column was avoided by the use of glass materials with the inner surfaces treated by high-temperature silylation. The day-to-day reproducibility of the method was represented by a relative standard deviation (RSD) of 5% for a felodipine concentration of 25 nmol/l. The minimum determinable concentration (giving better than 15% RSD) was 1-2 nmol/l (0.4-0.8 ng/ml).
血浆中的非洛地平用甲苯萃取,通过带有电子捕获检测的自动毛细管气相色谱法测定。通过使用内表面经高温硅烷化处理的玻璃材料,避免了二氢吡啶衍生物在进样器和色谱柱中发生不良氧化。该方法的日常重现性以非洛地平浓度为25 nmol/l时相对标准偏差(RSD)为5%表示。最低可测定浓度(RSD优于15%)为1 - 2 nmol/l(0.4 - 0.8 ng/ml)。