Hegde Sanjeet, Miquel Marc E, Boubertakh Redha, Gilderdale David, Muthurangu Vivek, Keevil Stephen F, Young Ian, Hill Derek L G, Razavi Reza S
Division of Imaging Sciences, King's College London School of Medicine, Guy's Campus, London SE1 9RT, UK.
J Vasc Interv Radiol. 2006 Jul;17(7):1175-9. doi: 10.1097/01.RVI.0000228466.09982.8B.
The lack of magnetic resonance (MR) safe catheters and guide wires remains an important obstacle to widespread clinical use of MR-guided endovascular procedures. The authors looked at the feasibility of using multiple tuned fiducial markers (TFM) and novel imaging sequences to track catheters reliably under MR and to evaluate the safety of such markers in terms of heating.
The visualization and tracking of a catheter with six quadrature tuned fiducial coils was carried out in a special designed in-vitro setup within a 1.5-T MR imager simulating an MR-guided endovascular intervention. The fiducial markers were also tested for heating.
The excellent signal contrast between the fiducial and the background when using novel interleaved real time and interactive sequences allowed for rapid and reliable identification of the fiducial markers and therefore the catheter. No significant heating of the marker was noted.
The authors have shown that catheters with multiple tuned fiducial markers are superior to passive catheter designs in terms of visualization and do not carry the risk of heating that is commonly associated with active catheters.
缺乏磁共振(MR)安全导管和导丝仍然是MR引导下血管内手术广泛临床应用的一个重要障碍。作者研究了使用多个调谐基准标记(TFM)和新型成像序列在MR下可靠跟踪导管的可行性,并从加热方面评估此类标记的安全性。
在1.5-T MR成像仪内的特殊设计体外装置中,使用六个正交调谐基准线圈对导管进行可视化和跟踪,模拟MR引导下的血管内介入。还对基准标记进行了加热测试。
当使用新型交错实时和交互式序列时,基准与背景之间具有出色的信号对比度,从而能够快速可靠地识别基准标记,进而识别导管。未观察到标记有明显发热。
作者表明,具有多个调谐基准标记的导管在可视化方面优于无源导管设计,并且不存在与有源导管通常相关的发热风险。