Suppr超能文献

人工耳蜗植入患者术后阻抗的变化:改良电极表面和鼓室内注射糖皮质激素的短期效果

Changes of postoperative impedances in cochlear implant patients: the short-term effects of modified electrode surfaces and intracochlear corticosteroids.

作者信息

Paasche Gerrit, Bockel Franziska, Tasche Claudia, Lesinski-Schiedat Anke, Lenarz Thomas

机构信息

Department of Otolaryngology, Medical University of Hannover, Hannover, Germany.

出版信息

Otol Neurotol. 2006 Aug;27(5):639-47. doi: 10.1097/01.mao.0000227662.88840.61.

Abstract

OBJECTIVE

The objective of this study was to investigate the effect of intraoperative application of steroid suspension and coating of the electrode contacts with a thin film of iridium oxide on the short-term, time-dependent development of the intracochlear impedance in adults implanted with the Nucleus 24 Contour electrode.

STUDY DESIGN

The time-dependent development of intracochlear impedances was investigated in four different groups of adult patients at daily and later weekly intervals until the first fitting. The four groups were as follows: 1) standard Nucleus 24 Contour (control, n = 7); 2) standard Nucleus 24 Contour with intraoperative application of steroids (Group S, n = 6); 3) iridium-coated Nucleus 24 Contour control (Group I, n = 8); and 4) iridium-coated Nucleus 24 Contour with intraoperative application of steroids (Group I + S, n = 5). All patients had postlinguistic onset of severe to profound sensorineural hearing loss and no or little benefit of conventional hearing aids. Absence of ossification or any other cochlear anomaly and also absence of signs of retrocochlear or central origin to the hearing impairment bilaterally had to be confirmed preoperatively.

RESULTS

Steroid application reduced impedances significantly (Groups S and I + S), whereas iridium coating lowered variance of the impedance among patients but did not reduce the impedance significantly. The steroid-induced reduction is more pronounced at basal electrode contacts. Furthermore, there is some indication that the tissue growth could be faster in patients having the iridium-coated Contour electrode.

CONCLUSION

Provided that the reduction of electrode impedances with application of steroids is persisting, intracochlear application of steroids can be considered on a regular basis. Iridium coating of the electrode contacts seems not to be justified to be included as standard procedure.

摘要

目的

本研究的目的是调查术中应用类固醇混悬液以及用氧化铱薄膜包覆电极触点对植入Nucleus 24 Contour电极的成人耳蜗内阻抗的短期、时间依赖性变化的影响。

研究设计

对四组成年患者的耳蜗内阻抗随时间的变化进行研究,在首次调试前每天及之后每周进行一次测量。四组如下:1)标准Nucleus 24 Contour(对照组,n = 7);2)术中应用类固醇的标准Nucleus 24 Contour(S组,n = 6);3)铱涂层Nucleus 24 Contour对照组(I组,n = 8);4)术中应用类固醇的铱涂层Nucleus 24 Contour(I + S组,n = 5)。所有患者均为语后发生的重度至极重度感音神经性听力损失,传统助听器无效果或效果甚微。术前必须确认无骨化或任何其他耳蜗异常,且双侧听力损失无蜗后或中枢性起源的迹象。

结果

应用类固醇显著降低了阻抗(S组和I + S组),而铱涂层降低了患者间阻抗的差异,但未显著降低阻抗。类固醇引起的阻抗降低在基底电极触点处更为明显。此外,有迹象表明,使用铱涂层Contour电极的患者组织生长可能更快。

结论

如果应用类固醇后电极阻抗的降低持续存在,则可考虑定期在耳蜗内应用类固醇。电极触点的铱涂层似乎没有理由作为标准程序纳入。