Comparison of vitros dry slide technology for determination of lithium ions with other methods.

作者信息

Serdarević Nafija, Malesić Ivan, Kozjek Franc

机构信息

Institute for Clinical Chemistry and Biochemistry, Sarajevo University Clinics Center, 71000 Sarajevo, Bolnicka 25, Bosnia and Herzegovina.

出版信息

Bosn J Basic Med Sci. 2006 May;6(2):32-6. doi: 10.17305/bjbms.2006.3191.

Abstract

The lithium ions concentration in human serum was determined using Dry-slide technology Vitros 250 Analyser (Ortho Clinical Diagnostic), atomic absorption spectrometry (AAS) method Perkin Elmer 403 and ion-selective electrode (ISE) potentiometry AVL 9181. We compared lithium ions results in sample sera between these methods. Our reference method was AAS. We analyzed lithium ions concentration in 23 sera samples of patients after oral administration of lithium carbonate (3x 300mg) Jadran, Galen Laboratory Rijeka, by dry-slide technology, AAS and ISE methods. The quality control, precision, reproducibility and accuracy for Vitros dry slide technology were assessed. We established that the main difference between AAS method and dry slide technology was not statistically significant at p< 0.05 according to Student t-test. Therefore, the dry slide technology may be a useful alternative or it may even replace other methods, such as AAS. The main difference between dry slide technology and ISE methods was statistically significant at p<0.05 using Student t-test. By ISE method, we obtained considerably higher results, which may be explained by the presence of electrolytes or medicaments interfering with lithium ions.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验