Crespi A, Bonsignore F, Paruccini N, Macchi I
U.O. Fisica Sanitaria, Azienda Ospedaliera S. Gerardo, Via Pergolesi 33, I-20052, Monza, Italy.
Radiol Med. 2006 Sep;111(6):863-75. doi: 10.1007/s11547-006-0081-9. Epub 2006 Aug 11.
The aim of this study was to calibrate monitors used in soft-copy review of diagnostic images in a pictures archiving and communication system (PACS) and to assess critical quality assurance (QA) parameters through appropriate checks.
Barco [cathode ray tube (CRT) and liquid crystal display (LCD)] and EIZO (LCD) monitors were evaluated. Calibration and QA controls were carried out during acceptance tests on the systems and every 6 months according to the Task Group 18 (TG18) report by the American Association of Physicists in Medicine (AAPM). The parameters in question include: maximum luminance, contrast ratio, luminance response, spatial resolution and angular response. A subjective evaluation of image quality was also conducted by a number of radiologists.
Barco medical monitors' results were well within tolerances, with significant parameters persisting over time. EIZO nonmedical monitors showed rapid performance deterioration below the minimum requirements. Calibration had to be repeated only in a few cases. Radiologists' evaluations showed that monitor quality is equal to or even better than that of conventional films.
Medical monitors turned out to be fully adequate to the task. Periodic QA tests are, however, absolutely necessary to ensure constant quality levels.
本研究的目的是校准图片存档与通信系统(PACS)中用于诊断图像软拷贝审阅的显示器,并通过适当检查评估关键质量保证(QA)参数。
对巴可(阴极射线管(CRT)和液晶显示器(LCD))和艺卓(LCD)显示器进行了评估。根据美国医学物理学家协会(AAPM)任务组18(TG18)报告,在系统验收测试期间以及每6个月进行一次校准和QA控制。所涉及的参数包括:最大亮度、对比度、亮度响应、空间分辨率和角度响应。还由多名放射科医生对图像质量进行了主观评估。
巴可医用显示器的结果完全在公差范围内,重要参数随时间持续存在。艺卓非医用显示器的性能迅速恶化至低于最低要求。仅在少数情况下需要重复校准。放射科医生的评估表明,显示器质量等于甚至优于传统胶片。
事实证明,医用显示器完全胜任这项任务。然而,定期进行QA测试对于确保质量水平恒定绝对必要。