Monitoring of dialysis water systems--is there a need for increased sampling?

作者信息

James R

机构信息

Dialysis Unit, Barts and The London NHS Trust, UK.

出版信息

EDTNA ERCA J. 2006 Apr-Jun;32(2):74-7. doi: 10.1111/j.1755-6686.2006.tb00454.x.

Abstract

Water quality is one of the most important aspects of ensuring a safe and effective delivery of haemodialysis. An extensive microbiological survey of several water systems highlighted a contamination problem which routine sampling had failed to detect. Current guidelines suggest that samples for microbiological and endotoxin analysis should be taken from the outlet of water treatment plant and points expected to have the highest bacterial load, normally the end of the distribution loop and connections to the dialysis machines, where the flow is at its lowest. The survey extended sampling to include a large number of machine connection points. Four systems were investigated. The samples from routine test points generally returned results within our operational limits (<10 CFU/ml and <0.06 IU/ml). However, results from several machine connection points exceeded these limits by a large margin. Several disinfection cycles were required in order to achieve results in keeping with our operational limits. The conclusion is that sample results from the end of the distribution loop may give a false sense of security by not indicating a contamination problem at the machine connection points. Increasing the number and frequency of machine connection points tested should provide greater security in detecting contamination and allow for remedial action at an earlier stage.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验