Chen Lih-Huei, Lai Po-Liang, Tai Ching-Lung, Niu Chi-Chien, Fu Tsai-Sheng, Chen Wen-Jer
Department of Orthopaedic Surgery, Chang Gung Memorial Hospital, Kweishan, Taoyuan, Taiwan.
Biomed Mater Eng. 2006;16(4):261-7.
Eight fresh porcine lumbar spines received a posterior instrumentation at L4-L5 using pedicle screw-rod system. Each specimen was tested utilizing laminectomies of varying extent. Group A (Integrity) preserved the spinous process and interspinous ligament; Group B (Partial laminectomy) removed the inferior portion of L4 spinous process and preserved the interspinous ligament of L3-L4; Group C (Complete laminectomy) removed the entire L4 spinous process. Hydraulic testing machine was used to generate an increasing moment up to 8400 N mm in flexion and extension. The intervertebral displacement on the superior adjacent disc between L3-L4 was measured using an extensometer. Under extension, no significant difference in the intervertebral displacement was observed among three different models of laminectomy. However, under flexion, the intervertebral displacement on adjacent disc with complete laminectomy was statistically larger than those of integrity and partial laminectomies (P=0.000976 and P=0.0363, respectively). No difference was found between integrity and partial laminectomy groups (P>0.05). This study implies that an instrumented spine with integrity of posterior complex is less likely to develop adjacent instability than a spine with destruction of the anchoring point for supraspinous ligament.
八具新鲜猪腰椎在L4-L5节段采用椎弓根螺钉-棒系统进行后路内固定。每个标本利用不同程度的椎板切除术进行测试。A组(完整组)保留棘突和棘间韧带;B组(部分椎板切除术组)切除L4棘突下部并保留L3-L4棘间韧带;C组(全椎板切除术组)切除整个L4棘突。使用液压试验机在屈伸过程中产生逐渐增加至8400 N·mm的力矩。使用引伸计测量L3-L4上相邻椎间盘的椎间位移。在伸展时,三种不同椎板切除模型之间未观察到椎间位移有显著差异。然而,在屈曲时,全椎板切除术组相邻椎间盘的椎间位移在统计学上大于完整组和部分椎板切除术组(分别为P=0.000976和P=0.0363)。完整组和部分椎板切除术组之间未发现差异(P>0.05)。本研究表明,后路复合体完整的内固定脊柱比棘上韧带锚固点破坏的脊柱发生相邻节段不稳定的可能性更小。