Is accurate rate response programming necessary?

作者信息

Sulke N, Dritsas A, Chambers J, Sowton E

机构信息

Department of Cardiology, Guy's Hospital, London, England.

出版信息

Pacing Clin Electrophysiol. 1990 Aug;13(8):1031-44. doi: 10.1111/j.1540-8159.1990.tb02150.x.

Abstract

Exercise capacity and general well-being are improved by appropriately programmed rate responsive pacemakers when compared to fixed rate units. Ten patients had activity sensing DDDR units implanted for combined AV block and sinus node incompetence. Ten patients had Sensolog activity sensing VVIR units implanted for complete heart block. The effects of over and under programming of rate response in both dual and single chamber activity sensor rate adaptive pacemakers has been assessed subjectively by visual analog scales and specific activity questionnaires and objectively by graded treadmill testing and the performance of standardized daily activities. Patients were randomly programmed to absent rate response (VVI in the Sensolog group), hyporesponsive (DDD in the dual chamber group), appropriate response (VVIR, DDDR according to Manufacturer's instructions) and over responsive (VVIR+, DDDR+) in a double-blind crossover design. Thirty percent of patients demanded early crossover from VVI, 30% from DDDR+ and 50% from VVIR+. Perception of Exercise Capability was similar to objective exercise treadmill times which were shorter in VVI than in VVIR or VVIR+ (P less than 0.05) or control subjects (P less than 0.001). There was no difference between any dual chamber mode or control subjects. General well-being was poorest in DDDR+ and VVIR+ modes despite objective improvement in exercise capacity. Symptoms were least in VVIR and DDDR and all but one patient chose appropriate programming as their overall preferred mode. Thus, even inaccurate rate response programming results in similar and improved exercise capacity compared to absent rate response but overprogramming is unacceptable to most patients, confirming that appropriate programming and sensor specificity is critical in rate responsive pacing.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验