Oosterkamp Barbara C M, van der Meer Wicher J, Rutenfrans Majelle, Dijkstra Pieter U
Department of Orthodontics, University Medical Centre Groningen, Groningen, the Netherlands.
Cleft Palate Craniofac J. 2006 Sep;43(5):519-23. doi: 10.1597/05-078.
To assess the reliability and validity of measurements performed on three-dimensional virtual models of neonatal bilateral cleft lip and palate patients, compared with measurements performed on plaster cast models.
Ten high-quality plaster cast models of bilateral cleft lip and palate patients were scanned with an LDI-scanner to obtain a three-dimensional virtual model. Linear measurements were performed on the plaster cast models using a digital caliper and also on the three-dimensional virtual model using Viscam RP version 2.1 software. The measurements were performed by two observers on two occasions.
Intraclass correlations ranging from .81 to .96 were found for all measurements except the measurement between the constructed reference point pr and reference point i (intraclass correlation = .40). A post hoc procedure in which top-view screen prints of the three-dimensional virtual model were used to perform the measurement between reference points pr and i demonstrated an intraclass coefficient of .90.
Three-dimensional virtual models obtained by laser scanning neonatal cast models of bilateral cleft lip and palate patients can be used reliably and validly to perform linear measurements between existing reference points on the surface of the model using Viscam RP version 2.1 software. Measurements between reference points constructed outside the surface of the model cannot be validly performed on the three-dimensional virtual model with the software used in this study. For these measurements, top-view screen prints of the three-dimensional virtual model can be used.
评估对新生儿双侧唇腭裂患者三维虚拟模型进行测量的可靠性和有效性,并与在石膏模型上进行的测量作比较。
使用LDI扫描仪对10个双侧唇腭裂患者的高质量石膏模型进行扫描,以获得三维虚拟模型。使用数字卡尺在石膏模型上进行线性测量,并使用Viscam RP 2.1版软件在三维虚拟模型上进行线性测量。测量由两名观察者分两次进行。
除构建的参考点pr和参考点i之间的测量(组内相关系数 = 0.40)外,所有测量的组内相关系数在0.81至0.96之间。使用三维虚拟模型的顶视图屏幕截图在参考点pr和i之间进行测量的事后程序显示组内系数为0.90。
通过激光扫描新生儿双侧唇腭裂患者的石膏模型获得的三维虚拟模型,可使用Viscam RP 2.1版软件在模型表面现有的参考点之间可靠且有效地进行线性测量。使用本研究中使用的软件,无法在三维虚拟模型上有效地进行在模型表面之外构建的参考点之间的测量。对于这些测量,可使用三维虚拟模型的顶视图屏幕截图。