9.4-T磁共振场中中耳和镫骨植入物的磁性特性。

Magnetic properties of middle ear and stapes implants in a 9.4-T magnetic resonance field.

作者信息

Fritsch Michael H, Gutt Jason J, Naumann Ilka C

机构信息

Department of Otolaryngology-Head and Neck Surgery, Indiana University Medical Center, Indianapolis, Indiana 46202, USA.

出版信息

Otol Neurotol. 2006 Dec;27(8):1064-9. doi: 10.1097/01.mao.0000233816.71671.b7.

Abstract

HYPOTHESIS

A 9.4-T magnetic resonance (MR) field may cause motion displacement of the middle ear and stapes implants not previously observed with 1.5- and 3.0-T magnets.

BACKGROUND

Publications have described the safety limitations of some otologic implants in 4.7-T field and resulted in several companywide patient safety-related recalls. To date, no studies have been reported for otologic implants in a 9.4-T MR field nor have comparisons been made with 4.7-T field strengths.

METHODS

Twenty-three commonly used middle ear and stapes prostheses were selected and exposed to 9.4-T MR fields in vitro within petri dishes, and eight of the 23 implants were further studied ex corpus in human temporal bones (TBs) in a 9.4-T MR field. This study has been approved by the institutional review board.

RESULTS

Eight prostheses in petri dishes grossly displaced at 9.4 T, three of which had not previously moved in either the 1.5- or 3.0-T magnets. The eight TB preparations showed no avulsions or motion indicators after exposure at 9.4 T.

CONCLUSION

Middle ear and stapes implants can move dramatically in petri dishes at 9.4-T MR field, more so than at 1.5 and 3.0 T. The absence of avulsions in the TB group strongly suggests that the surgical means used to fixate the middle ear implants to the middle ear structures successfully overcomes the magnetic moment produced at MR field strengths up to 9.4 T. The use of MR imaging is not contraindicated by this study's findings.

摘要

假设

9.4特斯拉的磁共振(MR)场可能会导致中耳和镫骨植入物发生运动位移,而这在之前1.5特斯拉和3.0特斯拉的磁体中未被观察到。

背景

已有文献描述了一些耳科植入物在4.7特斯拉场中的安全局限性,并导致了几次全公司范围内与患者安全相关的召回事件。迄今为止,尚无关于耳科植入物在9.4特斯拉MR场中的研究报告,也未与4.7特斯拉场强进行比较。

方法

选择23种常用的中耳和镫骨假体,在培养皿中于体外暴露于9.4特斯拉的MR场中,并且在23个植入物中的8个在人颞骨(TB)的离体状态下在9.4特斯拉的MR场中进行了进一步研究。本研究已获得机构审查委员会的批准。

结果

培养皿中的8个假体在9.4特斯拉时发生了明显位移,其中3个在之前1.5特斯拉或3.0特斯拉的磁体中均未移动。8个TB制剂在9.4特斯拉暴露后未显示撕脱或运动迹象。

结论

中耳和镫骨植入物在9.4特斯拉MR场的培养皿中会发生显著移动,比在1.5特斯拉和3.0特斯拉时更明显。TB组中未出现撕脱强烈表明,用于将中耳植入物固定到中耳结构的手术方法成功克服了高达9.4特斯拉MR场强产生的磁矩。本研究结果并不表明MR成像的使用是禁忌的。