Suppr超能文献

温度对根管封闭剂粘度的影响。

The effect of temperature on viscosity of root canal sealers.

作者信息

Lacey S, Pitt Ford T R, Yuan X-F, Sherriff M, Watson T

机构信息

Department of Biomaterials and Conservative Dentistry, Dental Institute, King's College London, London, UK.

出版信息

Int Endod J. 2006 Nov;39(11):860-6. doi: 10.1111/j.1365-2591.2006.01154.x.

Abstract

AIM

To test the hypothesis that there was no significant (alpha = 0.05) change in viscosity of commercially available root canal sealers with increase in temperature using a high-performance Advanced Rheometric Expansion System (ARES) rheometer.

METHODOLOGY

Materials tested were Apexit, Tubliseal EWT, Grossman's, AH Plus and Ketac-endo. Cone-and-plate geometry was used (25-mm diameter, 0.1 radian and gap 0.051 mm). Measurements were carried out for steady-state viscosity at 25 and 37 degrees C in the shear rate range of 0.001-50 s(-1) at standardized relative humidity and within 30 min from the start of mixing. Five samples were taken for each sealer at each temperature.

RESULTS

At 25 degrees C all sealers demonstrated shear thinning. At 37 degrees C Grossman's (powder : liquid ratio 2 : 1 and 3 : 1) and Ketac-endo had a rapid rise in viscosity and early set whereas the other sealers were shear thinning. On increasing temperature from 25 degrees C to 37 degrees C, Apexit, Tubliseal and AH Plus had reduced viscosity whereas Grossman's 2 : 1, Grossman's 3 : 1 and Ketac-endo had increased viscosity, which varied with the shear rate. The change in viscosity with change in temperature was significant (P < 0.05) for all sealers except AH Plus.

CONCLUSIONS

There was a variation in the effect of increasing temperature on each sealer depending on the shear rate. With the exception of AH Plus, a significant (P < 0.05) change in viscosity was found, and the null hypothesis was rejected.

摘要

目的

使用高性能的高级流变扩展系统(ARES)流变仪,检验关于市售根管封闭剂的粘度不会随温度升高而发生显著变化(α = 0.05)这一假设。

方法

所测试的材料为Apexit、Tubliseal EWT、Grossman氏封闭剂、AH Plus和Ketac-endo。采用锥板几何形状(直径25毫米、弧度0.1、间隙0.051毫米)。在标准相对湿度下且在混合开始后30分钟内,于25和37摄氏度下、在0.001 - 50 s(-1) 的剪切速率范围内测量稳态粘度。每种封闭剂在每个温度下取五个样本。

结果

在25摄氏度时,所有封闭剂均表现出剪切变稀。在37摄氏度时,Grossman氏封闭剂(粉液比2 : 1和3 : 1)和Ketac-endo的粘度迅速升高并早期凝固,而其他封闭剂则为剪切变稀。从25摄氏度升温至37摄氏度时,Apexit、Tubliseal和AH Plus的粘度降低,而Grossman氏2 : 1、Grossman氏3 : 1和Ketac-endo的粘度升高,且随剪切速率而变化。除AH Plus外,所有封闭剂的粘度随温度变化均具有显著性(P < 0.05)。

结论

根据剪切速率,温度升高对每种封闭剂的影响存在差异。除AH Plus外,发现粘度有显著(P < 0.05)变化,原假设被拒绝。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验