Suppr超能文献

保留交叉韧带型和后稳定型全膝关节置换术的股骨后滚:日常生活活动中的体内荧光透视分析

Femoral rollback of cruciate-retaining and posterior-stabilized total knee replacements: in vivo fluoroscopic analysis during activities of daily living.

作者信息

Fantozzi Silvia, Catani Fabio, Ensini Andrea, Leardini Alberto, Giannini Sandro

机构信息

Department of Electronics, Computer Science and Systems, Università di Bologna, Viale Risorgimento 2, 40136 Bologna, Italy.

出版信息

J Orthop Res. 2006 Dec;24(12):2222-9. doi: 10.1002/jor.20306.

Abstract

Restoration of the physiological flexor/extensor mechanism at the knee in terms of appropriate muscular lever arms, proper required quadriceps force, and suitable patellofemoral compressive force, is fundamental for the success of total knee replacement. Therefore, measurements of anteroposterior translation of the femoral component over the tibial base-plate against joint flexion during daily living activities are essential for the assessment of the in vivo performance of current prosthesis designs. Patients treated with posterior stabilized and cruciate retaining prostheses with excellent clinical scores were evaluated during stair climbing, sitting and rising from a chair, and step up and down, using a three-dimensional pose reconstruction technique based on videofluoroscopy. The posterior stabilized patients experienced a fairly consistent and physiological rollback specific of each motor task, demonstrating proper function of the spine-cam mechanism. Rollback was somehow inconsistent among subjects in the cruciate retaining group, accompanied with a smaller range of knee flexion. In this group, more posterior locations of the condyles correlated significantly with higher clinical and functional scores. Articular surface conformity restores physiological rollback in the presence of a spine-cam mechanism, but not coherently in the presence of the posterior cruciate ligament.

摘要

根据适当的肌肉力臂、所需的合适股四头肌力量以及合适的髌股关节压缩力,恢复膝关节的生理屈伸机制,是全膝关节置换成功的基础。因此,在日常生活活动中测量股骨部件相对于胫骨基板在关节屈曲时的前后平移,对于评估当前假体设计的体内性能至关重要。使用基于视频荧光透视的三维姿势重建技术,对临床评分优异的后稳定型和保留交叉韧带型假体治疗的患者在爬楼梯、从椅子上坐起和站起以及上下台阶时进行评估。后稳定型假体的患者在每个运动任务中都经历了相当一致且符合生理的后滚,表明棘突-凸轮机制功能正常。保留交叉韧带组的受试者之间后滚在某种程度上不一致,同时膝关节屈曲范围较小。在该组中,髁的更靠后位置与更高的临床和功能评分显著相关。关节面贴合度在存在棘突-凸轮机制时可恢复生理后滚,但在存在后交叉韧带时则不一致。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验