Sanchis-Alfonso Vicente
Department of Orthopaedics, Hospital Arnau de Vilanova, Valencia, Spain.
Knee Surg Sports Traumatol Arthrosc. 2007 Apr;15(4):361-4. doi: 10.1007/s00167-006-0207-6. Epub 2006 Oct 7.
We report on a case of a patient who had femoral osteolysis and severe metallosis of the knee joint resulting from the secondary abrasive wear of the metal-backed tibial component of a unicompartmental knee arthroplasty due to massive wear of tibial polyethylene. The failure was diagnosed 11 years after implantation. This metallosis affected the intra-articular tissues and the subchondral bone.
我们报告了一例患者,该患者因单髁膝关节置换术中金属背衬胫骨部件的胫骨聚乙烯大量磨损导致继发性磨蚀,出现了股骨骨质溶解和膝关节严重金属中毒。植入11年后诊断出失败情况。这种金属中毒影响了关节内组织和软骨下骨。