Lead interaction: rare cause of oversensing during implantation procedure of implantable cardioverter-defibrillator system.

作者信息

Gardas Rafal, Mlynarski Rafal, Staszak Katarzyna, Drzewiecka Anna, Pilat Eugeniusz, Zajac Tadeusz, Kargul Wlodzimierz

机构信息

Department of Electrocardiology, Medical University of Silesia, Katowice, Poland.

出版信息

Pacing Clin Electrophysiol. 2006 Oct;29(10):1174-5. doi: 10.1111/j.1540-8159.2006.00510.x.

Abstract

A variety of etiologies can cause erroneous detection in patients with implantable cardioverter defibrillator (ICDs). Interaction between two endocardial leads is rare and uncommon in causing electrical noise. During a reimplantation procedure of an ICD system in a 68-year-old man, additional electrical signals could be detected. The interaction between two endocardial defibrillation leads was identified as the cause of sensing problems. When it is not possible to extract the nonfunctional endocardial lead during implantation of the new electrode, it should be implanted away and not in parallel from the old one to avoid interaction between them. A variety of etiologies can cause erroneous detection in patiens with ICD. Interaction between two endocardial leads is rare and uncommon to cause electrical noise.