Gardas Rafal, Mlynarski Rafal, Staszak Katarzyna, Drzewiecka Anna, Pilat Eugeniusz, Zajac Tadeusz, Kargul Wlodzimierz
Department of Electrocardiology, Medical University of Silesia, Katowice, Poland.
Pacing Clin Electrophysiol. 2006 Oct;29(10):1174-5. doi: 10.1111/j.1540-8159.2006.00510.x.
A variety of etiologies can cause erroneous detection in patients with implantable cardioverter defibrillator (ICDs). Interaction between two endocardial leads is rare and uncommon in causing electrical noise. During a reimplantation procedure of an ICD system in a 68-year-old man, additional electrical signals could be detected. The interaction between two endocardial defibrillation leads was identified as the cause of sensing problems. When it is not possible to extract the nonfunctional endocardial lead during implantation of the new electrode, it should be implanted away and not in parallel from the old one to avoid interaction between them. A variety of etiologies can cause erroneous detection in patiens with ICD. Interaction between two endocardial leads is rare and uncommon to cause electrical noise.