静息和运动期间多导联QT间期计算机测量的可重复性。

Reproducibility of computerized measurements of QT interval from multiple leads at rest and during exercise.

作者信息

Hekkala Anna-Mari, Väänänen Heikki, Swan Heikki, Oikarinen Lasse, Viitasalo Matti, Toivonen Lauri

机构信息

Helsinki University Hospital, Department of Cardiology, Helsinki, Finland.

出版信息

Ann Noninvasive Electrocardiol. 2006 Oct;11(4):318-26. doi: 10.1111/j.1542-474X.2006.00124.x.

Abstract

BACKGROUND

Accurate measurement of the QT interval is important for diagnosing long QT syndrome (LQTS), and in research on determinants of ventricular repolarization time. We tested automatic analysis of QT intervals from multiple ECG leads on chest.

METHODS

Eleven healthy volunteers and 10 genotyped LQTS patients were tested at rest and during exercise with a bicycle ergometer twice 1-31 months apart. Electrocardiograms were recorded with the body surface potential mapping system, and 12 precordial channels were selected for analysis. Averaged QT peak and QT end intervals were determined with an automated algorithm, and the difference QT end minus QT peak (Tp-e) was calculated. Repeatability was assessed by coefficient of variation (CV) between measurements.

RESULTS

Within one test at rest the QT end intervals were highly repeatable with CV 0.6%. In repeated tests CV was 4.4% for QT end interval and 3.5% when the QT interval was corrected for heart rate. In exercise test at specified heart rates, mean CV was 3.0% for QT end and 2.9% for QT peak interval. The CV of Tp-e interval was 10.2% at rest, and 9.3% in exercise test. Reproducibility was comparable between healthy subjects and LQTS patients.

CONCLUSIONS

The BSPM system with automated analysis produced accurate and highly repeatable QT interval measurements. Reproducibility was adequate also over prolonged time periods both at rest and in exercise stress test. The method can be applied in studying duration of ventricular repolarization time in different physiologic and pharmacologic interventions.

摘要

背景

准确测量QT间期对于诊断长QT综合征(LQTS)以及心室复极时间决定因素的研究至关重要。我们测试了从胸部多个心电图导联自动分析QT间期。

方法

11名健康志愿者和10名基因分型的LQTS患者在静息状态和使用自行车测力计进行运动时接受测试,两次测试间隔1 - 31个月。使用体表电位标测系统记录心电图,并选择12个胸前导联进行分析。采用自动算法确定平均QT峰间期和QT终末间期,并计算QT终末间期减去QT峰间期的差值(Tp - e)。通过测量之间的变异系数(CV)评估重复性。

结果

在一次静息测试中,QT终末间期具有高度重复性,CV为0.6%。在重复测试中,QT终末间期的CV为4.4%,心率校正后的QT间期CV为3.5%。在指定心率的运动测试中,QT终末间期的平均CV为3.0%,QT峰间期为2.9%。Tp - e间期在静息时的CV为10.2%,运动测试中为9.3%。健康受试者和LQTS患者之间的可重复性相当。

结论

具有自动分析功能的体表电位标测系统产生了准确且高度可重复的QT间期测量结果。在静息和运动应激测试中,长时间内的可重复性也足够。该方法可应用于研究不同生理和药物干预下心室复极时间的持续时间。

相似文献

7
The QT and corrected QT interval in recovery after exercise in children.儿童运动后恢复期的 QT 和校正 QT 间期。
Circ Arrhythm Electrophysiol. 2011 Aug;4(4):448-55. doi: 10.1161/CIRCEP.110.961094. Epub 2011 May 23.

本文引用的文献

3
Molecular basis of congenital and acquired long QT syndromes.先天性和获得性长QT综合征的分子基础。
J Electrocardiol. 2004;37 Suppl:1-6. doi: 10.1016/j.jelectrocard.2004.08.002.
4
The measurement of the Q-T interval of the electrocardiogram.心电图Q-T间期的测量。
Circulation. 1952 Sep;6(3):378-88. doi: 10.1161/01.cir.6.3.378.
7
More light on QT interval measurement.QT间期测量的更多见解。
Heart. 2002 Mar;87(3):193-4. doi: 10.1136/heart.87.3.193.
8
Cellular mechanisms underlying the long QT syndrome.长QT综合征的细胞机制。
Curr Opin Cardiol. 2002 Jan;17(1):43-51. doi: 10.1097/00001573-200201000-00007.
10
T peak-Tend interval as an index of transmural dispersion of repolarization.
Eur J Clin Invest. 2001 Jul;31(7):555-7. doi: 10.1046/j.1365-2362.2001.00849.x.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验