Suppr超能文献

同位素稀释气相色谱/质谱法中测量不确定度的预测

Prediction of measurement uncertainty in isotope dilution gas chromatography/mass spectrometry.

作者信息

Hasegawa Hiroshi, Shinohara Yoshihiko, Hashimoto Takao, Matsuda Rieko, Hayashi Yuzuru

机构信息

School of Pharmacy, Tokyo University of Pharmacy and Life Science, 1432-1 Horinouchi, Hachioji, Tokyo 192-0392, Japan.

出版信息

J Chromatogr A. 2006 Dec 15;1136(2):226-30. doi: 10.1016/j.chroma.2006.09.073. Epub 2006 Oct 11.

Abstract

An equation is theoretically derived which describes the relative standard deviation (RSD) of the amount ratios of analyte to its isotope-labeled variant in gas chromatography/mass spectrometry (GC/MS) using the stable isotope dilution method. The determination of methyltestosterone is taken as an example. The uncertainty equation proposed is justified by comparing the theoretical RSD values with the experimental RSD values obtained by replication over a wide range of analyte amount. The detection limit and quantitation limit are estimated from the continuous plot (precision profile) of the theoretical RSD against analyte amount.

摘要

从理论上推导了一个方程,该方程描述了在使用稳定同位素稀释法的气相色谱/质谱联用(GC/MS)中,分析物与其同位素标记变体的量比的相对标准偏差(RSD)。以甲基睾酮的测定为例。通过将理论RSD值与在宽范围分析物量下通过重复实验获得的实验RSD值进行比较,验证了所提出的不确定度方程。根据理论RSD相对于分析物量的连续图(精密度曲线)估算检测限和定量限。