Prochaczek F, Galecka J
Silesian Medical Academy, First Medical Department, Katowice and Medical Electronic Research Center, Zabrze, Poland.
Pacing Clin Electrophysiol. 1990 Dec;13(12 Pt 2):2022-5. doi: 10.1111/j.1540-8159.1990.tb06935.x.
The quality of the ECG recording during transcutaneous pacing was evaluated in six healthy volunteers. The transcutaneous pacing stimulator was an NP4D special unit to which was attached an electronic suppressor of artifact generated by the transcutaneous stimulating impulse. The relationship between this suppression of artifact distortion and the resulting QRS complex was analyzed. The results revealed that the suppression time (described as a 2-mm oscillation from 100-Hz frequency) was required for ECG distortion elimination and that this is dependent on the threshold of ventricular pacing. The width of the resulting QRS complex diminishes as the suppression is extended over 110 ms. These results suggest the necessity of individual adjustment of the suppression time so that the efficacy of transcutaneous pacing is adequately assessed.
在六名健康志愿者身上评估了经皮起搏期间心电图记录的质量。经皮起搏刺激器是一个NP4D特殊装置,其上连接了一个由经皮刺激脉冲产生的伪差电子抑制器。分析了这种伪差失真抑制与由此产生的QRS波群之间的关系。结果显示,消除心电图失真需要抑制时间(描述为100赫兹频率下2毫米的振荡),并且这取决于心室起搏阈值。随着抑制时间延长超过110毫秒,由此产生的QRS波群宽度会减小。这些结果表明需要对抑制时间进行个体化调整,以便充分评估经皮起搏的效果。