Marshall N W
Clinical Physics Group, Barts and the London NHS Trust, St Bartholomew's Hospital, London, EC1A 6BE, UK.
Phys Med Biol. 2006 Nov 7;51(21):5655-73. doi: 10.1088/0031-9155/51/21/018. Epub 2006 Oct 16.
This paper describes objective and subjective image quality measurements acquired as part of a routine quality assurance (QA) programme for an amorphous selenium (a-Se) full field digital mammography (FFDM) system between August-04 and February-05. During this period, the FFDM detector developed a fault and was replaced. A retrospective analysis of objective image quality parameters (modulation transfer function (MTF), normalized noise power spectrum (NNPS) and detective quantum efficiency (DQE)) is presented to try and gain a deeper understanding of the detector problem that occurred. These measurements are discussed in conjunction with routine contrast-detail (c-d) results acquired with the CDMAM (Artinis, The Netherlands) test object. There was significant reduction in MTF over this period of time indicating an increase in blurring occurring within the a-Se converter layer. This blurring was not isotropic, being greater in the data line direction (left to right across the detector) than in the gate line direction (chest wall to nipple). The initial value of the 50% MTF point was 6 mm(-1); for the faulty detector the 50% MTF points occurred at 3.4 mm(-1) and 1.0 mm(-1) in the gate line and data line directions, respectively. Prior to NNPS estimation, variance images were formed of the detector flat field images. Spatial distribution of variance was not uniform, suggesting that the physical blurring process was not constant across the detector. This change in variance with image position implied that the stationarity of the noise statistics within the image was limited and that care would be needed when performing objective measurements. The NNPS measurements confirmed the results found for the MTF, with a strong reduction in NNPS as a function of spatial frequency. This reduction was far more severe in the data line direction. A somewhat tentative DQE estimate was made; in the gate line direction there was little change in DQE up to 2.5 mm(-1) but at the Nyquist frequency the DQE had fallen to approximately 35% of the original value. There was severe attenuation of DQE in the data line direction, the DQE falling to less than 0.01 above approximately 3.0 mm(-1). C-d results showed an increase in threshold contrast of approximately 25% for details less than 0.2 mm in diameter, while no reduction in c-d performance was found at the largest detail diameters (1.0 mm and above). Despite the detector fault, the c-d curve was found to pass the European protocol acceptable c-d curve.
本文描述了在2004年8月至2005年2月期间,作为非晶硒(a-Se)全视野数字乳腺摄影(FFDM)系统常规质量保证(QA)计划一部分所获取的客观和主观图像质量测量结果。在此期间,FFDM探测器出现故障并被更换。本文对客观图像质量参数(调制传递函数(MTF)、归一化噪声功率谱(NNPS)和探测量子效率(DQE))进行了回顾性分析,以试图更深入地了解所发生的探测器问题。这些测量结果与使用CDMAM(荷兰阿廷尼斯公司)测试物体获得的常规对比度-细节(c-d)结果相结合进行讨论。在此时间段内,MTF显著降低,表明a-Se转换层内的模糊增加。这种模糊不是各向同性的,在数据线方向(探测器从左到右)比在栅极线方向(胸壁到乳头)更严重。50%MTF点的初始值为6mm⁻¹;对于有故障的探测器,50%MTF点在栅极线和数据线方向分别出现在3.4mm⁻¹和1.0mm⁻¹处。在估计NNPS之前,由探测器平场图像形成方差图像。方差的空间分布不均匀,这表明整个探测器上的物理模糊过程并非恒定。方差随图像位置的这种变化意味着图像内噪声统计的平稳性有限,并且在进行客观测量时需要谨慎。NNPS测量结果证实了MTF的测量结果,NNPS随空间频率有显著降低。这种降低在数据线方向更为严重。进行了一个有点初步的DQE估计;在栅极线方向,直到2.5mm⁻¹,DQE几乎没有变化,但在奈奎斯特频率处,DQE已降至原始值的约35%。在数据线方向,DQE严重衰减,在约3.0mm⁻¹以上,DQE降至小于0.01。c-d结果显示,对于直径小于0.2mm 的细节,阈值对比度增加了约25%,而在最大细节直径(1.0mm及以上)处未发现c-d性能下降。尽管探测器出现故障,但发现c-d曲线通过了欧洲协议可接受的c-d曲线。