Johnston S L, Bradding P, Watkins J
Department of Cardiology, St. Mary's Hospital, Portsmouth Hampshire, England.
Pacing Clin Electrophysiol. 1991 Jan;14(1):20-7. doi: 10.1111/j.1540-8159.1991.tb04043.x.
We compared the rate response to low level work, arm exercise, step testing, and treadmill exercise between the Siemens Sensolog 703 S (Sg) and the Medtronic Activitrax (Ax) pacemakers, and simultaneous sinus rhythm (SR) in an elderly population. In ten subjects mean age 70.6 years, range 49-81 years, the pacemaker responses were noninvasively compared by strapping the units to the chest wall under constant and equal pressure applied by an adjustable belt. Pacemaker rates and SR were recorded simultaneously on a three-channel ECG recorder. The units were programmed in situ to give an increase in rate from 70 beats/min sitting, to 100 beats/min after walking at a normal pace for 30 seconds. Programming took three times longer for Sg (P less than 0.02). The response to standing and bending was poor for both units (Ax mean 75.7 beats/min, Sg mean 77.6 beats/min), when compared to SR (mean 90.3 beats/min). A 20-step test resulted in a greater rate response from Ax (mean 105.3 beats/min) than from Sg (mean 96.3 beats/min [P = 0.09]), though both were still less than SR (116.3). There were significant differences between the two pacemakers on treadmill testing, at peak rate achieved (Ax mean 112.5 beats/min; Sg mean 102.7 beats/min [P less than 0.005], SR mean 122.5 beats/min) and at end exercise (Ax mean 112.5 beats/min; Sg mean 92.9 beats/min [P less than 0.002], SR mean 121.3 beats/min). Arm exercise, however, resulted in a significantly greater response from Sg than Ax (105.1 and 88.5 beats/min, respectively; P less than 0.01, SR 98.7 beats/min).(ABSTRACT TRUNCATED AT 250 WORDS)
我们比较了西门子Sensolog 703 S(Sg)和美敦力Activitrax(Ax)起搏器以及老年人群中的同时性窦性心律(SR)对低强度工作、手臂运动、阶梯试验和跑步机运动的心率反应。在10名平均年龄70.6岁(范围49 - 81岁)的受试者中,通过用可调节腰带在胸壁上施加恒定且相等的压力将起搏器固定,对起搏器反应进行无创比较。在三通道心电图记录仪上同时记录起搏器心率和SR。将这些装置在原位编程,使其心率从坐着时的70次/分钟增加到以正常步伐行走30秒后的100次/分钟。Sg的编程时间长了三倍(P小于0.02)。与SR(平均90.3次/分钟)相比,两个装置对站立和弯腰的反应都较差(Ax平均75.7次/分钟,Sg平均77.6次/分钟)。20步试验中,Ax的心率反应(平均105.3次/分钟)大于Sg(平均96.3次/分钟[P = 0.09]),但两者仍低于SR(116.3)。在跑步机测试中,两个起搏器在达到的峰值心率(Ax平均112.5次/分钟;Sg平均102.7次/分钟[P小于0.005],SR平均122.5次/分钟)和运动结束时(Ax平均112.5次/分钟;Sg平均92.9次/分钟[P小于0.002],SR平均121.3次/分钟)存在显著差异。然而,手臂运动导致Sg的反应明显大于Ax(分别为105.1和88.5次/分钟;P小于0.01,SR为98.7次/分钟)。(摘要截短于250字)