Suppr超能文献

基于浓缩染料的激光扫描共聚焦显微镜强度校准

Intensity calibration of a laser scanning confocal microscope based on concentrated dyes.

作者信息

Model Michael A, Blank James L

机构信息

Department of Biological Sciences, Kent State University, Kent, Ohio, USA.

出版信息

Anal Quant Cytol Histol. 2006 Oct;28(5):253-61.

Abstract

OBJECTIVE

To find water-soluble fluorescent dyes with absorption in various regions of the spectrum and investigate their utility as standards for laser scanning confocal microscopy.

STUDY DESIGN

Several dyes were found to have characteristics required for fluorescence microscopy standards. The intensity of biological fluorescent specimens was measured against the emission of concentrated dyes. Results using different optics and different microscopes were compared.

RESULTS

Slides based on concentrated dyes can be prepared in a highly reproducible manner and are stable under laser scanning. Normalized fluorescence of biological specimens remains consistent with different objective lenses and is tolerant to some mismatch in optical filters or imperfect pinhole alignment. Careful choice of scanning parameters is necessary to ensure linearity of intensity measurements.

CONCLUSION

Concentrated dyes provide a robust and inexpensive intensity standard that can be used in basic research or clinical studies.

摘要

目的

寻找在光谱各区域有吸收的水溶性荧光染料,并研究其作为激光扫描共聚焦显微镜标准物的效用。

研究设计

发现几种染料具有荧光显微镜标准所需的特性。根据浓缩染料的发射来测量生物荧光标本的强度。比较了使用不同光学器件和不同显微镜的结果。

结果

基于浓缩染料的载玻片可以以高度可重复的方式制备,并且在激光扫描下稳定。生物标本的归一化荧光在使用不同物镜时保持一致,并且对滤光片的一些不匹配或针孔对准不完善具有耐受性。需要仔细选择扫描参数以确保强度测量的线性。

结论

浓缩染料提供了一种强大且廉价的强度标准,可用于基础研究或临床研究。