Device interaction--antitachycardia pacemakers and defibrillators for sustained ventricular tachycardia.

作者信息

Ahern T S, Nydegger C, McCormick D J, Marinchak R, Kowey P, Horowitz L N, Worley S, Kutalek S P

机构信息

Philadelphia Arrhythmia Group, Hahnemann University, PA 19102.

出版信息

Pacing Clin Electrophysiol. 1991 Feb;14(2 Pt 2):302-7. doi: 10.1111/j.1540-8159.1991.tb05111.x.

Abstract

We evaluated the combined use of permanent automatic antitachycardia pacemakers and implanted defibrillators in ten patients with recurrent monomorphic sustained ventricular tachycardia (VT). Pacemaker programming was VVI-T automatic burst in eight patients, VVI-T magnet mode in one patient, and VVI in one patient. Device interactions occurred in four patients, requiring changes in pacemaker programming. These included defibrillator multiple counting during pacing, inappropriate pacemaker bursts initiating VT, inappropriate reset of the pacemaker antitachycardia mode by defibrillation, defibrillator discharge after pacemaker VT termination, and defibrillator VT reinitiation. Two patients required pacemaker programming out of the antitachycardia mode, and two required a change in antitachycardia pacing parameters. Seven patients remain in automatic VVI-T and three in VVI modes. Mean follow-up is 13 months and all patients are alive. Thus, although pacemaker/defibrillator combinations function well for patients with more than one VT rate, device interactions occur frequently and may require pacemaker reprogramming or elimination of the overdrive mode. Combined use of these devices should be cautiously considered when single device therapy is unsatisfactory. Devices that combine both pacing and defibrillation features may reduce adverse interaction.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验