使用心包外电极板和植入式除颤器电极对犬进行除颤后心脏病理变化的差异。

Differences in the pathological changes in dogs' hearts after defibrillation with extrapericardial paddles and implanted defibrillator electrodes.

作者信息

Harman D, Matthews R, Gilli N, McCulloch R

机构信息

Telectronics Pacing Systems, Lane Cove, New South Wales, Australia.

出版信息

Pacing Clin Electrophysiol. 1991 Feb;14(2 Pt 2):358-61. doi: 10.1111/j.1540-8159.1991.tb05121.x.

Abstract

A comparison was made between the pathological changes in the myocardium of eight dogs, each receiving about 90 joules of energy in a series of defibrillation discharges, delivered either between paddles placed against the pericardium (3 dogs) or between implanted Telectronics 040-105 defibrillation patch electrodes (5 dogs). The changes in the myocardium were most pronounced where the paddles had been applied to the pericardium. There was transmural damage beneath the left and right paddle positions and in the surrounding tissues. Extensive subepicardial and subendocardial myocyte damage was obvious histologically in the right ventricle of one of the patch dogs and in all of the paddle dogs. The percentage of damaged myocardial mass, both right ventricular and total involvement, was higher in the three paddle dogs compared with the five patch dogs. There was septal damage in the heart of one paddle dog. Necrosis of the right ventricular wall was observed in three of the patch dogs and in all the three paddle dogs. Scattered necrotic myocytes and some patches of mild necrosis up to 1-mm deep were observed in the left ventricle of the patch dogs (severity score 1-3). The necrosis was more extensive in the paddle dogs, ranging from mild necrosis less than 1-mm deep to marked necrosis incorporating half-to-whole ventricular wall thickness (severity score 3-5).

摘要

对八只狗的心肌病理变化进行了比较,每只狗在一系列除颤放电中接受约90焦耳的能量,除颤放电通过置于心包上的电极板(3只狗)或植入的Telectronics 040 - 105除颤贴片电极(5只狗)进行。心肌变化在电极板与心包接触的部位最为明显。在左右电极板位置下方及周围组织存在透壁损伤。在一只贴片狗的右心室以及所有电极板狗的右心室中,组织学上可见广泛的心外膜下和心内膜下心肌细胞损伤。与五只贴片狗相比,三只电极板狗的受损心肌质量百分比,无论是右心室还是整体受累情况,都更高。一只电极板狗的心脏出现了室间隔损伤。在三只贴片狗和所有三只电极板狗中均观察到右心室壁坏死。在贴片狗的左心室中观察到散在的坏死心肌细胞和一些深度达1毫米的轻度坏死灶(严重程度评分为1 - 3)。电极板狗的坏死范围更广,从深度小于1毫米的轻度坏死到累及心室壁厚度一半至整个厚度的明显坏死(严重程度评分为3 - 5)。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验