Liang Jian-hui, Zhou Xing, Peng Pin-xian, Zheng Zhao-bin, Yu Xiao-li
Department of Chest Surgery, The Second Affiliated Hospital of Guangzhou Medical College, Guangzhou 510260, China.
Zhonghua Wai Ke Za Zhi. 2006 Jul 15;44(14):952-5.
To observe the feasibility of a segment thoracic esophagus replaced by using a nitinol alloy composite artificial esophagus.
A nitinol alloy composite artificial esophagus was made of a nitinol tube coated with a silicone membrane with a dacron flange 1 cm proximal to each end. There were ten pigs for the models of experimental animal. The models of experimental animal were established by excised a segment of 7 cm long thoracic esophagus and replaced with the graft. In accordance with the different anastomosis connective way 10 pigs were divided into group I and group II, every group had 5 pigs. In group I monolayer end-to-end anastomosis was used with this prosthesis. In group II monolayer end-to-end anastomosis was used with this prosthesis and above the site of anastomosis adding dacron strip fixed. After operation the operate animals were observed for eating and health. Esophagography and flexible esophagoscopy were performed once a month at the first, second, and third postoperatively month. Three pigs of the grafted prostheses sloughed out spontaneously at different postoperatively period were killed for autopsy and pathologic examination of the neoesophagus.
All pigs had no leakage at the site of anastomosis and survived over 6 months. The longest survival was 20 months. Seven of 10 pigs the grafted prostheses sloughed out spontaneously at 2 to 6 months after the operation (4 in group I and 3 in group II). The grafted prostheses sloughed out spontaneously less than 3 months after operation, marked neoesophageal stenosis in the grafted portion was recognized (3 in group I). They displayed serious difficulty in eating. But the prostheses sloughed out spontaneously more than 3 months after operation, neoesophageal stenosis was increased slightly and the stenosis was only temporary and gradually resolved. (1 in group I, 3 in group II) They lived in good health and without difficulty in eating. Three of 10 pigs the grafted prosthesis still stayed in the grafted place over 10 months (1 in group I, 2 in group II). The neoesophageal had no stenosis. They were perfectly well in eating.
A segment excised pig esophagus replaced with a nitinol alloy composite artificial esophagus is feasible and takes a better result.
观察采用镍钛合金复合人工食管替代胸段食管的可行性。
镍钛合金复合人工食管由镍钛管制成,其外覆硅胶膜,两端各1 cm处有涤纶法兰。选用10头猪作为实验动物模型。通过切除7 cm长的胸段食管并植入移植物建立实验动物模型。根据不同的吻合连接方式将10头猪分为Ⅰ组和Ⅱ组,每组5头。Ⅰ组采用单层端端吻合方式使用该假体。Ⅱ组采用单层端端吻合方式使用该假体,并在吻合部位上方加用涤纶条固定。术后观察实验动物的进食和健康状况。术后第1、2、3个月每月进行一次食管造影和纤维食管镜检查。对术后不同时期自发脱落的3头植入假体的猪进行处死,对新食管进行尸检和病理检查。
所有猪吻合部位均无渗漏,存活超过6个月。最长存活时间为20个月。10头猪中有7头在术后2至6个月假体自发脱落(Ⅰ组4头,Ⅱ组3头)。术后3个月内假体自发脱落,吻合部位出现明显的新食管狭窄(Ⅰ组3头)。它们进食困难严重。但术后3个月后假体自发脱落,新食管狭窄略有增加,且狭窄仅为暂时性,逐渐缓解。(Ⅰ组1头,Ⅱ组3头)它们健康状况良好,进食无困难。10头猪中有3头植入的假体在植入部位停留超过10个月(Ⅰ组1头,Ⅱ组2头)。新食管无狭窄。它们进食完全正常。
采用镍钛合金复合人工食管替代切除的猪胸段食管是可行的,且效果较好。