Suppr超能文献

胫骨柄设计对骨水泥型全膝关节置换术后骨密度的影响:一项前瞻性七年随访研究

Influence of the tibial stem design on bone density after cemented total knee arthroplasty: a prospective seven-year follow-up study.

作者信息

Hernandez-Vaquero Daniel, Garcia-Sandoval Manuel A, Fernandez-Carreira Jose M, Gava Richard

机构信息

School of Medicine, University of Oviedo, Oviedo, Spain.

出版信息

Int Orthop. 2008 Feb;32(1):47-51. doi: 10.1007/s00264-006-0280-y. Epub 2006 Nov 18.

Abstract

We prospectively measured the changes in bone mineral density (BMD) in the proximal tibia of 20 total knee arthroplasties, ten with cruciform stems and ten with cylindrical stems. The measurements were made one, four and seven years after surgery. We observed a uniform density decrease in three regions of interest from one to seven years of follow-up. Cylindrical stems showed an asymmetrical density decrease between the three regions of interest, with no change in the central region, a slight decrease in the lateral region, and large decrease in the medial region. Multivariate analysis with general linear model showed the stem type factor as statistically significant for medial region of interest (p = 0.006). The cylindrical stem produces heterogeneous BMD changes under the tibial platform in knee arthroplasties, and this could be a potential risk factor for asymmetrical subsidence of this component.

摘要

我们前瞻性地测量了20例全膝关节置换术中胫骨近端的骨密度(BMD)变化,其中10例使用十字形柄,10例使用圆柱形柄。测量在术后1年、4年和7年进行。我们观察到在随访的1至7年中,三个感兴趣区域的密度均呈均匀下降。圆柱形柄在三个感兴趣区域之间显示出不对称的密度下降,中央区域无变化,外侧区域略有下降,内侧区域大幅下降。采用一般线性模型的多变量分析显示,柄类型因素在感兴趣的内侧区域具有统计学意义(p = 0.006)。在膝关节置换术中,圆柱形柄在胫骨平台下会产生不均匀的骨密度变化,这可能是该组件不对称下沉的潜在风险因素。

相似文献

引用本文的文献

本文引用的文献

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验