Suppr超能文献

多孔钛气管假体的研制:一项在绵羊身上的研究。

Development of tracheal prostheses made of porous titanium: a study on sheep.

作者信息

Schultz P, Vautier D, Charpiot A, Lavalle P, Debry C

机构信息

Department of Otorhinolaryngology and Technical Research Team, CHU Hautepierre and ERT 10-61, Strasbourg Cedex, France.

出版信息

Eur Arch Otorhinolaryngol. 2007 Apr;264(4):433-8. doi: 10.1007/s00405-006-0195-7. Epub 2006 Nov 23.

Abstract

Authors report the development of a biomaterial to be used for tracheal and laryngeal reconstruction. This experimentation follows the replacement of trachea in rats with porous titanium implants. The aim of the study is to test this type of prosthesis on sheep, whose trachea is of comparable size to that of humans. Six ewes were implanted with porous titanium implants after resection of 5 cm of trachea. The planned period for the implantation was from 3 to 6 months before the sacrifice of the animals for histological analysis. After a simple immediate postoperative course, the implantations developed complications of tracheal patency, responsible for four deaths (tracheal obstruction by mucous plug n = 2, inferior necrosis of trachea n = 1, pneumopathy n = 1). The two remaining sheep presented no complications. The mechanical performance of the prostheses was good. The histological results showed an inflammatory stenosis of the tracheo-prosthetic junctions, which was not the direct result of death. The protheses were integrated by the surrounding tissue, but endoprosthetic colonisation by pseudostratified ciliated columnar epithelium was low or nil. The absence of endoprosthetic lining was responsible for the complications. The biocompatibility of the biomaterial is not in question, but the surgical procedure will have to be modified by an endoprosthetic mucous graft before implantation so as to accelerate healing process.

摘要

作者报告了一种用于气管和喉部重建的生物材料的研发情况。该实验是在大鼠身上用多孔钛植入物替换气管之后进行的。本研究的目的是在绵羊身上测试这种类型的假体,绵羊的气管大小与人类的相当。6只母羊在切除5厘米气管后植入了多孔钛植入物。计划的植入期为3至6个月,之后处死动物进行组织学分析。术后初期过程简单,但植入物出现了气管通畅方面的并发症,导致4只羊死亡(黏液栓引起气管阻塞n = 2,气管下部坏死n = 1,肺病n = 1)。剩下的2只羊未出现并发症。假体的机械性能良好。组织学结果显示气管-假体连接处存在炎性狭窄,这并非导致死亡的直接原因。假体被周围组织整合,但假复层纤毛柱状上皮对假体内部的定植程度很低或没有。假体内部缺乏衬里是导致并发症的原因。生物材料的生物相容性没有问题,但在植入前必须通过假体内部黏液移植来改进手术程序,以加速愈合过程。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验