Dual-contrast single breath-hold 3D abdominal MR imaging.

作者信息

Winkelmann Richard, Börnert Peter, De Becker Jan, Hoogeveen Romhild, Mazurkewitz Peter, Dössel Olaf

机构信息

Institute of Biomedical Engineering, University of Karlsruhe, 76128, Karlsruhe, Germany.

出版信息

MAGMA. 2006 Dec;19(6):297-304. doi: 10.1007/s10334-006-0057-8. Epub 2006 Nov 24.

Abstract

OBJECT

Multiple contrasts are often helpful for a comprehensive diagnosis. In 3D abdominal MRI, breath-hold techniques are preferred for single contrast acquisitions to avoid respiratory artifacts. In this paper, highly accelerated parallel MRI is used to acquire large 3D abdominal volumes with two different contrasts within a single breath-hold.

MATERIAL AND METHODS

In vivo studies have been performed on six healthy volunteers, combining T (1)- and T (2)-weighted, gradient- or spin-echo based scans, as well as water/fat resolved imaging in a single breath-hold. These 3D scans were acquired with an acceleration factor of six, using a prototype 32-element receive array.

RESULTS

The presented approach was tested successfully on all volunteers. The whole liver area was covered by a FOV of 350 x 250 x 200 mm(3) for all scans with reasonable spatial resolution. Arbitrary scan protocols generating different contrasts have been shown to be combinable in this single breath-hold approach. Good spatial correspondence with negligible spatial offset was achieved for all different scan combinations acquired in overall breath-hold times between 15 and 25 s.

CONCLUSION

Enabled by highly parallel imaging technology, this study demonstrates the technical feasibility and the promising image quality of single breath-hold dual contrast MRI.