The purpose of this study is to assess the criterion validity of peroneal and posterior tibial nerve conduction measurements obtained with the NC-stat system. Sixty patients referred to the Boston VA EMG laboratory were enrolled. Each subject had a full study of the lower extremity performed using traditional EMG equipment prior to obtaining the NC-stat measurements. These included peroneal and posterior tibial distal motor latency (DML), amplitude (AMP) and F-wave latency (FLAT) measurements. Excellent criterion validity was demonstrated for the peroneal and posterior tibial FLATs and the peroneal AMP. Acceptable criterion validity was identified in the peroneal DML and the posterior tibial AMP. The validity of the posterior tibial DML could not be demonstrated. With the exception of the peroneal DML, criterion validity was maintained in a sub-group analysis of the 50% most abnormal parameter values. The comparability of NCS performed with the NC-stat and in traditional settings has been demonstrated for motor studies of the median and ulnar nerves in previous studies. This study shows that the technology used by the NC-stat for studying the peroneal and posterior tibial nerves compares favorably as well with that obtained with traditional EMG equipment used under neurologist supervision.
本研究的目的是评估使用NC-stat系统获得的腓总神经和胫后神经传导测量的标准效度。招募了60名转诊至波士顿退伍军人事务部肌电图实验室的患者。在获得NC-stat测量结果之前,使用传统肌电图设备对每个受试者的下肢进行了全面检查。这些检查包括腓总神经和胫后神经的远端运动潜伏期(DML)、波幅(AMP)和F波潜伏期(FLAT)测量。结果表明,腓总神经和胫后神经的F波潜伏期以及腓总神经的波幅具有出色的标准效度。在腓总神经DML和胫后神经波幅方面确定了可接受的标准效度。胫后神经DML的效度无法得到证实。除腓总神经DML外,在50%最异常参数值的亚组分析中标准效度得以维持。先前的研究已证明,在正中神经和尺神经的运动研究中,使用NC-stat进行的神经传导检查(NCS)与传统检查方法具有可比性。本研究表明,NC-stat用于研究腓总神经和胫后神经的技术与在神经科医生监督下使用传统肌电图设备获得的技术相比也毫不逊色。