Abraham L M, Epasinghe N C R, Selva D, Casson R
Department of Ophthalmology and Visual Sciences, South Australian Institute of Ophthalmology, Royal Adelaide Hospital, Adelaide, South Australia, Australia.
Eye (Lond). 2008 Apr;22(4):503-6. doi: 10.1038/sj.eye.6702669. Epub 2006 Dec 8.
To assess the agreement between ICare rebound tonometer and Goldmann applanation tonometer in the hands of experienced and inexperienced tonometrists.
Two tonometrists, experienced with both Goldmann applanation tonometry (GAT) and ICare Tonometry (ICT) measured intraocular pressure (IOP), in a masked fashion, in 100 patients. In another series of 58 patients, ICT was performed by an inexperienced tonometrist and GAT by an experienced tonometrist.
In approximately 80% of patients, the difference in IOP between GAT and ICT was <or=2 mmHg in group 1 and <or=3 mmHg in group 2. The 95% limits of agreement were -4.0-4.4 mmHg in group 1 and -6.0-5.0 mmHg in group 2.
ICT compares reasonably with GAT, in both experienced and inexperienced hands. Its ease of use, portability, and sterility make it an attractive tonometer. Its degree of accuracy in inexperienced hands would make it a useful instrument for health care workers with limited ophthalmic experience.