Helander A, Beck O, Wennberg M, Wikström T, Jacobsson G
Department of Psychiatry, Karolinska Institute, St. Göran's Hospital, Stockholm, Sweden.
Anal Biochem. 1991 Jul;196(1):170-3. doi: 10.1016/0003-2697(91)90134-f.
A method for the routine quantitative determination of the major serotonin metabolite 5-hydroxyindole-3-acetic acid (5-HIAA) in urine is described. 5-HIAA was analyzed without prior sample cleanup, using an automated high-performance liquid chromatography system with isocratic elution and electrochemical detection (+0.60 V versus a Ag/AgCl reference electrode). The urine samples were mixed with a solution of the internal standard (5-hydroxyindole-3-propionic acid) and centrifuged. The supernatant was transferred to sealed glass vials, and a 2-microliters aliquot was injected directly onto a C18 reversed-phase analytical column, using an automatic sample injector. Samples of urine could be stored for several months at -80 or at +7 degrees C for 2 days without loss of 5-HIAA. However, a gradual decline with time occurred in crude samples stored at room temperature or above, as well as in urine samples diluted with the mobile phase. The detector response was linear in the range of 0-65 mumol/l 5-HIAA, and the intra- and interassay coefficients of variation were about 5 and 7%, respectively (n = 10).
本文描述了一种常规定量测定尿液中主要血清素代谢物5-羟吲哚-3-乙酸(5-HIAA)的方法。使用具有等度洗脱和电化学检测(相对于Ag/AgCl参比电极+0.60 V)的自动高效液相色谱系统,无需事先对样品进行净化即可对5-HIAA进行分析。将尿液样品与内标(5-羟吲哚-3-丙酸)溶液混合并离心。将上清液转移至密封玻璃小瓶中,并使用自动进样器将2微升等分试样直接注入C18反相分析柱。尿液样品可在-80℃下保存数月,或在+7℃下保存2天而不会损失5-HIAA。然而,在室温或更高温度下储存的粗样品以及用流动相稀释的尿液样品中,5-HIAA会随时间逐渐下降。检测器响应在0-65μmol/l 5-HIAA范围内呈线性,批内和批间变异系数分别约为5%和7%(n = 10)。