Housden R James, Gee Andrew H, Treece Graham M, Prager Richard W
University of Cambridge, Department of Engineering, Trumpington Street, Cambridge CB2 1PZ, UK.
Ultrasound Med Biol. 2006 Dec;32(12):1897-904. doi: 10.1016/j.ultrasmedbio.2006.06.027.
Freehand 3D ultrasound can be acquired without a position sensor by deducing the elevational probe motion from the interframe speckle decorrelation. However, a freehand scan involves lateral and axial, as well as elevational, probe motion. The lateral sampling is determined by the A-line separation and is relatively sparse: lateral motion tracking therefore requires subsample interpolation. In this paper, we investigate the resilience of lateral interpolation techniques to simultaneous lateral and elevational probe motion. We propose a novel interpolation strategy and, through a series of in vitro experiments, compare its performance with that of established alternatives. The new technique is shown to be superior, limiting interpolation errors to around 5% of the length of the freehand reconstruction.
通过从帧间散斑去相关推导仰角探头运动,徒手三维超声无需位置传感器即可获取。然而,徒手扫描涉及探头的横向、轴向以及仰角运动。横向采样由A线间距决定,且相对稀疏:因此横向运动跟踪需要子样本插值。在本文中,我们研究了横向插值技术对探头同时进行横向和仰角运动的适应能力。我们提出了一种新颖的插值策略,并通过一系列体外实验将其性能与现有替代方法进行比较。结果表明,新技术更具优势,可将插值误差限制在徒手重建长度的5%左右。