用色度计和色度视觉系统测量天然牙颜色的比较。

Comparison of the color of natural teeth measured by a colorimeter and Shade Vision System.

作者信息

Cho Byeong-Hoon, Lim Yong-Kyu, Lee Yong-Keun

机构信息

Department of Conservative Dentistry and Dental Research Institute, College of Dentistry, Seoul National University, Seoul, Republic of Korea.

出版信息

Dent Mater. 2007 Oct;23(10):1307-12. doi: 10.1016/j.dental.2006.11.008. Epub 2006 Dec 18.

Abstract

OBJECTIVES

The objectives were to measure the difference in the color and color parameters of natural teeth measured by a tristimulus colorimeter (CM, used as a reference) and Shade Vision System (SV), and to determine the influence of color parameters on the color difference between the values measured by two instruments.

METHOD

Color of 12 maxillary and mandibular anterior teeth was measured by CM and SV for 47 volunteers (number of teeth=564). Color parameters such as CIE L*, a* and b* values, chroma and hue angle measured by two instruments were compared. Chroma was calculated as Cab=(a2 = b2)1/2, and hue angle was calculated as h degrees =arctan(b/a*). The influence of color parameters measured by CM on the color difference (DeltaE*(ab)) between the values measured by two instruments was analyzed with multiple regression analysis (alpha=0.01).

RESULTS

Mean DeltaE*(ab) value between the values measured by two instruments was 21.7 (+/-3.7), and the mean difference in lightness (CIE L*) and chroma was 16.2 (+/-3.9) and 13.2 (+/-3.0), respectively. Difference in hue angle was high as 132.7 (+/-53.3) degrees . Except for the hue angle, all the color parameters showed significant correlations and the coefficient of determination (r(2)) was in the range of 0.089-0.478. Based on multiple regression analysis, the standardized partial correlation coefficient (beta) of the included predictors for the color difference was -0.710 for CIE L* and -0.300 for C*(ab) (p<0.01).

CONCLUSION

All the color parameters showed significant but weak correlations except for hue angle. When lightness and chroma of teeth were high, color difference between the values measured by two instruments was small. Clinical accuracy of two instruments should be investigated further.

摘要

目的

本研究旨在测量用三刺激色度计(CM,作为参考)和色度视觉系统(SV)测量的天然牙颜色及颜色参数的差异,并确定颜色参数对两种仪器测量值之间颜色差异的影响。

方法

对47名志愿者的12颗上颌和下颌前牙进行颜色测量,使用CM和SV(牙齿数量 = 564颗)。比较两种仪器测量的颜色参数,如CIE L*、a和b值、色度和色相角。色度计算为Cab = (a2 + b2)1/2,色相角计算为h度 = arctan(b/a*)。用多元回归分析(α = 0.01)分析CM测量的颜色参数对两种仪器测量值之间颜色差异(ΔE*(ab))的影响。

结果

两种仪器测量值之间的平均ΔE*(ab)值为21.7(±3.7),明度(CIE L*)和色度的平均差异分别为16.2(±3.9)和13.2(±3.0)。色相角差异高达132.7(±53.3)度。除色相角外,所有颜色参数均显示出显著相关性,决定系数(r(2))在0.089 - 0.478范围内。基于多元回归分析,CIE L的标准化偏相关系数(β)为 - 0.710,C(ab)为 - 0.300(p < 0.01)。

结论

除色相角外,所有颜色参数均显示出显著但较弱的相关性。当牙齿的明度和色度较高时,两种仪器测量值之间的颜色差异较小。两种仪器的临床准确性应进一步研究。