Suppr超能文献

现代陶瓷全髋关节置换术中陶瓷内衬的延迟性骨折

Delayed fracture of a ceramic insert with modern ceramic total hip replacement.

作者信息

Min Byung-Woo, Song Kwang-Soon, Kang Chul-Hyung, Bae Ki-Cheol, Won Ye-Yeon, Lee Kwon-Yong

机构信息

Department of Orthopedic Surgery, School of Medicine, Keimyung University, Daegu, South Korea.

出版信息

J Arthroplasty. 2007 Jan;22(1):136-9. doi: 10.1016/j.arth.2005.12.012. Epub 2006 Jun 21.

Abstract

We report a case of ceramic liner fracture in a 34-year-old man that occurred 1 year 9 months after total hip arthroplasty with a BiCONTACT (Aesculap, Tuttlingen, Germany) ceramic-on-ceramic prosthesis. The ceramic liner was found within the metal shell in 4 large pieces and multiple small fragments in the peripheral portions. Its condition indicated that the possible mechanism of late ceramic liner chip fracture was repeated episodes of impingement between the prosthetic neck and the edge of the ceramic liner. Squatting, kneeling, and sitting cross-legged caused impingement. These positions are more common in Asian populations than in Western populations. After the joint was thoroughly irrigated, a new modular ceramic liner and a 28-mm ceramic head were implanted.

摘要

我们报告了一例34岁男性患者陶瓷内衬骨折的病例,该骨折发生在使用BiCONTACT(德国图特林根蛇牌)陶瓷对陶瓷假体进行全髋关节置换术后1年9个月。陶瓷内衬在金属壳内被发现有4个大块碎片以及周边部分的多个小碎片。其情况表明,陶瓷内衬晚期碎片骨折的可能机制是假体颈部与陶瓷内衬边缘之间反复发生撞击。深蹲、跪姿和盘腿坐会导致撞击。这些姿势在亚洲人群中比在西方人群中更常见。在关节彻底冲洗后,植入了一个新的模块化陶瓷内衬和一个28毫米的陶瓷股骨头。