Suppr超能文献

[五个拉丁美洲国家的放射学服务质量]

[The quality of radiology services in five Latin American countries].

作者信息

Fleitas Ileana, Caspani Carlos C, Borrás Cari, Plazas Maria-Cristina, Miranda Alberto A, Brandan Maria-Ester, de la Mora Roxana

机构信息

Organización Panamericana de la Salud, Washington, DC 20037, USA.

出版信息

Rev Panam Salud Publica. 2006 Aug-Sep;20(2-3):113-24. doi: 10.1590/s1020-49892006000800008.

Abstract

OBJECTIVE

To determine the correlation between certain quality indicators for imaging services and the accurate interpretation of radiological exams for four frequent complaints: breast lumps, gastrointestinal discomfort, back pain, and symptoms of tuberculosis.

METHODS

Twenty-six radiology services in Argentina, Bolivia, Colombia, Cuba, and Mexico were assessed. The mammography and conventional radiographic/fluoroscopic equipment used in selected services were evaluated utilizing common protocols, data sheets, testing instruments, phantoms, and calibrated dosimetry systems. The studies were performed in medium-complexity facilities. Informed consent was obtained from all patients studied, and the confidentiality of results was guaranteed. The following parameters were documented: type of facility (public vs. private); population covered; patient workload; radiological and image-processing equipment and supplies; education and training of professional and technical staff; quality assurance and preventive maintenance programs, and adherence to radiation safety standards. The performance of x-ray units, image receptors and processors; darkroom and image viewing conditions; patient doses and image quality, were determined using standardized parameters in all cases. Independent panels of radiologists, recognized as experts by the local radiological society, assessed the quality of the clinical images obtained and performed a radiological interpretation for each patient using the same films and clinical history available to the institution's imaging physicians. The agreement between the panel of expert's reports and those of local radiologists was taken as an indicator of the radiological diagnostic accuracy.

RESULTS

Analyses were carried out of 366 mammograms, 343 radiological procedures for gastrointestinal complaints, 319 X-rays of the spinal column, and 157 chest radiographs. The agreement between the radiological interpretation of the panel of experts and of the local physician ranged from 70% to 100%, except in the case of spinal column films in Cuba (57.8%) and of mammograms in Mexico (33.3%), which the panel of experts found to be among those having the poorest quality. There was a significant positive correlation between the accuracy of the radiological interpretation and the quality of the radiological images. Image quality showed a positive correlation with the technicians' level of education and training. Studies performed in services that had automatic film processors and that complied with the indicators established for screen-film contact yielded better images and a higher proportion of studies with concordant results. More than 50% of the viewboxes did not satisfy the quality criteria for luminance and homogeneity.

CONCLUSIONS

A good quality image is critical to achieving an accurate diagnosis. Emphasis should be placed on the continuing education of radiology technicians and on the acquisition and maintenance of adequate equipment and accessories, especially viewboxes, intensifying screens, and automatic film processors, given the impact they have on image quality.

摘要

目的

确定影像服务的某些质量指标与针对四种常见病症(乳腺肿块、胃肠道不适、背痛和结核病症状)的放射学检查准确解读之间的相关性。

方法

对阿根廷、玻利维亚、哥伦比亚、古巴和墨西哥的26家放射科进行了评估。使用通用方案、数据表、测试仪器、模体和校准剂量测定系统,对选定科室使用的乳腺X线摄影和传统放射/荧光透视设备进行了评估。研究在中等复杂度的机构中进行。获得了所有研究患者的知情同意,并保证了结果的保密性。记录了以下参数:机构类型(公立与私立);覆盖人群;患者工作量;放射学和图像处理设备及用品;专业和技术人员的教育与培训;质量保证和预防性维护计划,以及对辐射安全标准的遵守情况。在所有情况下,均使用标准化参数确定X射线设备、图像接收器和处理器的性能;暗室和图像查看条件;患者剂量和图像质量。由当地放射学会认可的放射科医生独立小组,评估所获得临床图像的质量,并使用该机构影像科医生所拥有的相同胶片和临床病史,对每位患者进行放射学解读。专家小组报告与当地放射科医生报告之间的一致性,被用作放射学诊断准确性的指标。

结果

对366例乳腺X线摄影、343例针对胃肠道不适的放射学检查、319例脊柱X线片和157例胸部X线片进行了分析。专家小组与当地医生的放射学解读之间的一致性在70%至100%之间,但古巴的脊柱X线片(57.8%)和墨西哥的乳腺X线摄影(33.3%)除外,专家小组认为这些是质量最差的。放射学解读的准确性与放射学图像的质量之间存在显著正相关。图像质量与技术人员的教育和培训水平呈正相关。在配备自动洗片机且符合屏-片接触所设定指标的科室进行的研究,获得了更好的图像,且结果一致的研究比例更高。超过50%的观片灯不满足亮度和均匀性的质量标准。

结论

高质量图像对于实现准确诊断至关重要。鉴于放射学技术人员的继续教育以及适当设备和配件(尤其是观片灯、增感屏和自动洗片机)的购置与维护对图像质量的影响,应予以重视。