A double-beam rapid-wavelength-scanning stopped-flow spectrophotometer system based on the Norcon model 501 spectrometer was construced, which enables u.v.-or visible absorbance spectra to be recorded at the rate of 800/s after the rapid mixing (within 3ms) of two reactant solutions. Each spectrum spans about 200nm in 1ms. It is possible to record difference spectra during reactions with half-lives less than 10ms involving absorbance changes of less than 0.1 absorbance unit. Analogue circuitry is used to produce spectra of absorbance against wavelength. Up to 32 such spectra can be recorded at pre-selected times during a reaction and stored in an 8Kx8-bit-word hard-wired data-capture system to be subsequently displaned individually or simultaneously. Time-courses at different wavelengths can also be displayed. By averaging up to 216 spectra it is possible to record spectra under conditions of low signal-to-noise ratios...
构建了基于Norcon 501型光谱仪的双光束快速波长扫描停流分光光度计系统,该系统能够在两种反应物溶液快速混合(3毫秒内)后以每秒800次的速率记录紫外或可见吸收光谱。每个光谱在1毫秒内跨越约200纳米。对于半衰期小于10毫秒且吸光度变化小于0.1吸光度单位的反应,可以记录差示光谱。模拟电路用于产生吸光度对波长的光谱。在反应过程中,可以在预先选定的时间记录多达32个这样的光谱,并存储在一个8K×8位字的硬连线数据采集系统中,随后可单独或同时显示。也可以显示不同波长下的时间进程。通过对多达216个光谱求平均,可以在低信噪比条件下记录光谱……