Suppr超能文献

使用遥测技术同时评估结肠腔内压力和通过时间。

Simultaneous assessment of the intraluminal pressure and transit time of the colon using a telemetry technique.

作者信息

Zhang W Q, Yan G Z, Ye D D, Chen C W

机构信息

Institute of Medical Precise Instrumentation, Shanghai Jiaotong University, Shanghai, PR China.

出版信息

Physiol Meas. 2007 Feb;28(2):141-8. doi: 10.1088/0967-3334/28/2/003. Epub 2006 Dec 12.

Abstract

Colonic motility disorders are common conditions. However, our understanding of normal and pathological motor functions of the colon remains limited, mainly due to the technical difficulties in accessing this organ for study. To investigate colonic motility under normal physiological conditions, we have developed a novel monitoring system based on a telemetry technique. The system is capable of prolonged and noninvasive measurement of intraluminal pressure changes and transit time of intra-colonic contents. To test the in vivo performance of the monitoring system, 15 healthy volunteers and 15 patients with functional constipation (FC) participated in this study. A single-use telemetry capsule embedded with sensors was ingested by the subjects. The capsule is capable of transmitting colonic pressure and temperature wirelessly. The time of the telemetry capsule entering the segmental colon was detected by ultrasonic detection of the batteries in capsule. Pressure recordings confirmed in general a circadian behavior of colonic motility, as well as its response to waking and meals. In the FC patients, the contractile response to morning awakening and meal ingestion was significantly lower compared to the controls. The transit time measured using this method agreed with the time calculated from radiopaque markers (r = 0.89, p < 0.05). The clinical study proved both the reliability and the noninvasiveness of the system. This capsule-style manometric system may represent a useful tool for the study of physiology and pathology of colonic motor disorders.

摘要

结肠动力障碍是常见病症。然而,我们对结肠正常和病理运动功能的理解仍然有限,主要是因为研究该器官存在技术困难。为了研究正常生理条件下的结肠动力,我们开发了一种基于遥测技术的新型监测系统。该系统能够长时间且无创地测量腔内压力变化和结肠内内容物的通过时间。为了测试监测系统的体内性能,15名健康志愿者和15名功能性便秘(FC)患者参与了本研究。受试者摄入一枚嵌入传感器的一次性遥测胶囊。该胶囊能够无线传输结肠压力和温度。通过超声检测胶囊中的电池来确定遥测胶囊进入结肠节段的时间。压力记录总体上证实了结肠动力的昼夜节律行为及其对醒来和进食的反应。与对照组相比,FC患者对早晨醒来和进食的收缩反应明显较低。用这种方法测得的通过时间与根据不透X线标志物计算出的时间一致(r = 0.89,p < 0.05)。临床研究证明了该系统的可靠性和无创性。这种胶囊式测压系统可能是研究结肠动力障碍生理和病理的有用工具。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验