Guan Y H, Smulders J, Fisher D, Sutherland I A
Brunel Institute for Bioengineering, Brunel University West London, Uxbridge, Middlesex UB8 3PH, UK.
J Chromatogr A. 2007 Jun 1;1151(1-2):115-20. doi: 10.1016/j.chroma.2006.12.097. Epub 2007 Jan 8.
Retention properties of polyethylene glycol-phosphate aqueous two-phase systems in a spiral coil (5 mm I.D.) on Type-J synchronous counter-current chromatographic devices have been compared for the elution mode where the lower phase is the mobile phase and flows from the inside head terminal. This was achieved with the aid of digital imaging under stroboscopic illumination, an image analysis and measurement of the displaced volume of the stationary phase. For the spiral coil, high and stable stationary phase retention at mobile phase flow rates up to 64 ml/min has been obtained. Wave-like disturbance of the interface near the proximal point was observed and analyses have been made for possible use in protein separation.