金属组织标记夹引起的伪影对乳腺MRI的影响。

Effects on breast MRI of artifacts caused by metallic tissue marker clips.

作者信息

Genson Charles C, Blane Caroline E, Helvie Mark A, Waits Seth A, Chenevert Thomas L

机构信息

Department of Radiology, University of Michigan Health System, B1 D520 UH, 1500 E Medical Center Dr., Ann Arbor, MI 48109-0030, USA.

出版信息

AJR Am J Roentgenol. 2007 Feb;188(2):372-6. doi: 10.2214/AJR.05.1254.

Abstract

OBJECTIVE

The purpose of our study was to investigate MR artifacts related to tissue marker clips used in breast imaging procedures.

MATERIALS AND METHODS

Breast phantoms were created using gelatin doped with gadolinium. Four commercially available tissue marker clips were evaluated. Clinical MR evaluation of all phantoms with 1.5-T gradient-recalled echo sequences was performed. Images were evaluated for size and character of the visible artifacts and graphically appreciable fat saturation inhomogeneities. Quantitative measurement of the local inhomogeneity in 3D parts per million maps was obtained as a function of distance from each tissue marker.

RESULTS

All tissue marker clips caused signal void artifacts on non-fat-suppressed images that measured 2-6 times the clip diameter. The degree of fat suppression inhomogeneity was minor but clinically appreciable. The local clip-induced field inhomogeneity varied from 0.25 to greater than 4.0 PPM for the four clips. At 0.25 PPM, the zonal diameter of frequency shift varied from 6 mm to 44 mm.

CONCLUSION

Artifacts caused by tissue marker clips could limit the sensitivity of MRI for detection and follow-up of breast cancer. The local effects on field inhomogeneity will affect local fat suppression and make spectroscopy data less reliable. These effects, though small, are measurable and vary among the different clips evaluated.

摘要

目的

我们研究的目的是调查乳腺成像程序中使用的组织标记夹相关的磁共振伪影。

材料与方法

使用掺钆明胶制作乳腺模型。评估了四种市售的组织标记夹。使用1.5-T梯度回波序列对所有模型进行临床磁共振评估。评估图像中可见伪影的大小和特征以及图形上可察觉的脂肪抑制不均匀性。获取三维百万分之一图中局部不均匀性的定量测量值,作为距每个组织标记距离的函数。

结果

所有组织标记夹在非脂肪抑制图像上均导致信号缺失伪影,其大小为夹子直径的2至6倍。脂肪抑制不均匀程度较小,但在临床上可察觉。四种夹子引起的局部场不均匀性从0.25到大于4.0百万分之一。在0.25百万分之一时,频移的带状直径从6毫米到44毫米不等。

结论

组织标记夹引起的伪影可能会限制磁共振成像对乳腺癌检测和随访的敏感性。对场不均匀性的局部影响将影响局部脂肪抑制,并使光谱数据的可靠性降低。这些影响虽然很小,但可测量,且在所评估的不同夹子之间有所不同。