Feke G T
Schepens Retina Associates Foundation for Clinical Research, Boston, MA 02215, USA.
Bull Soc Belge Ophtalmol. 2006(302):171-84.
The theory underlying the development of laser Doppler instrumentation for the measurement of retinal blood flow is framed in terms of (a) the enunciation of the Doppler principle; (b) the invention of the laser; and (c) the invention of the technique known as optical mixing spectroscopy. The features of the instrumentation, beginning with the first prototype in 1972 and culminating with the introduction of the Canon Laser Blood Flowmeter in 1998 are presented in detail. Results from seven separate studies reporting on the reproducibility of retinal blood flow measurements using the Canon instrument, as well as a review of 12 separate presentations made at the 2004 annual meeting of the Association for Research in Vision and Ophthalmology (ARVO) using the Canon instrument in studies involving retinal circulatory physiology and associated clinical research are also presented.
用于测量视网膜血流的激光多普勒仪器的发展所依据的理论,是根据以下几点构建的:(a)多普勒原理的阐明;(b)激光的发明;以及(c)被称为光学混合光谱学的技术的发明。详细介绍了该仪器的特点,从1972年的第一个原型开始,到1998年佳能激光血流仪的推出达到顶峰。还展示了七项独立研究的结果,这些研究报告了使用佳能仪器测量视网膜血流的可重复性,以及对2004年视觉与眼科学研究协会(ARVO)年会上使用佳能仪器进行的12项独立报告的综述,这些报告涉及视网膜循环生理学及相关临床研究。