[输液泵与患者安全]

[Infusion pumps and patient safety].

作者信息

Bjørn Brian, Garde Kim, Pedersen Beth Lilja

机构信息

Hvidovre Hospital, Enhed for Patientsikkerhed, Afsnit 423.

出版信息

Ugeskr Laeger. 2007 Jan 22;169(4):315-8.

Abstract

INTRODUCTION

Infusion pumps are important clinical tools where controlled and precise infusions are needed. However, there are a number of potential risks for patient safety in their use: There is a risk of free-flow, i.e. an uncontrolled infusion that may have fatal consequences to the patient; staff is often poorly trained to use these devices, and often the devices themselves do not offer much support in this regard.

MATERIAL AND METHODS

Technical and organisational data regarding the use of infusion pumps in the hospitals of the Copenhagen Hospital Corporation (CHC) were collected and analysed. Critical incidents reported to the CHC Incident Database were analysed.

RESULTS

Forty-two different infusion pumps are in use by the five CHC hospitals. There was a total of 919 volumetric infusion pumps, 71% having set-based free-flow protection. Critical incidents were in 40% of cases caused by user-error, the most common being setting an incorrect infusion rate. Discontinuation of the infusion was reported in 27% of incident reports, the causes being disconnection or kinking of the infusion line; this resulted in two cases of awareness under anaesthesia.

CONCLUSION

Centralising the decision process for procurement and establishing a central library of standardised equipment might well reduce patient safety risks and any under-use of equipment. Usability testing prior to procurement decisions is vital; such testing should be performed by validated methods and not simply by putting the pumps to test in a clinical setting. More knowledge about user-friendly designs of medical equipment is needed.

摘要

引言

输液泵是需要进行精确控制输注的重要临床工具。然而,在其使用过程中存在一些对患者安全的潜在风险:存在自由流动的风险,即不受控制的输注,这可能对患者造成致命后果;工作人员通常未接受良好的使用这些设备的培训,而且这些设备本身在这方面往往也提供不了太多支持。

材料与方法

收集并分析了哥本哈根医院集团(CHC)各医院使用输液泵的技术和组织数据。对报告给CHC事件数据库的严重事件进行了分析。

结果

CHC的五家医院共使用42种不同的输液泵。共有919台容量输液泵,其中71%具备基于设置的自由流动保护功能。40%的严重事件是由用户错误导致的,最常见的是设置了错误的输注速率。27%的事件报告中出现了输液中断,原因是输液管断开或扭结;这导致了两例麻醉下知晓的情况。

结论

集中采购决策过程并建立标准化设备中央库很可能会降低患者安全风险以及设备的任何未充分使用情况。采购决策前的可用性测试至关重要;此类测试应通过经过验证的方法进行,而不仅仅是在临床环境中对泵进行测试。需要更多关于医疗设备用户友好设计的知识。