爱普生Expression 1680平板扫描仪与维达尔VXR-16剂量测定专业胶片扫描仪在使用放射变色胶片和射线照相胶片进行调强放射治疗剂量测定中的比较。

Comparison of the epson expression 1680 flatbed and the vidar VXR-16 dosimetry PRO film scanners for use in IMRT dosimetry using gafchromic and radiographic film.

作者信息

Wilcox Ellen, Daskalov George, Nedialkova Lucy

机构信息

Radiation Oncology, Saint Francis Hospital and Medical Center, 114 Woodland Street, Hartford, Connecticut 06105, USA.

出版信息

Med Phys. 2007 Jan;34(1):41-8. doi: 10.1118/1.2402584.

Abstract

Intensity-modulated radiotherapy (IMRT) treatment plan verification is often done using Kodak EDR2 film and a Vidar Dosimetry PRO film digitizer. However, since many hospitals are moving towards a filmless environment, access to a film processor may not be available. Therefore, we have investigated a newly available Gafchromic EBT film for IMRT dosimetry. Planar IMRT dose distributions are delivered to both EBT and EDR2 film and scanned with the Vidar VXR-16 as well as an Epson Expression 1680 flatbed scanner. The measured dose distributions are then compared to those calculated with a Pinnacle treatment planning system. The IMRT treatments consisted of 7-9 6 MV beams for treatment of prostate, head and neck, and a few other sites. The films were analyzed using FilmQATM (3cognition LLC) software. Comparisons between measured and calculated dose distributions are reported as dose difference (DD) (pixels within +/-5%), distance to agreement (DTA) (3 mm), as well as gamma values (y) (dose= +/-3%, dist. =2 mm). Using EDR2 with the Vidar scanner is an established technique and agreement between calculated and measured dose distributions was better than 90% in all indices (DD, DTA, and gamma). However, agreement with calculations deteriorated reaching the lower 80% for EBT film scans with the Vidar scanner in logarithmic mode. The EBT Vidar scans obtained in linear mode showed an improved agreement to the upper 80% range, but artifacts were still observed across the scan. These artifacts were very distinct in all EBT scans and can be attributed to the way the film is transported through the scanner. In the Epson scanner both films are rigidly immobilized and the light source scans over the film. It was found that the Epson scanner performed equally well with both types of film giving agreement to better than 90% in all indices.

摘要

调强放射治疗(IMRT)治疗计划验证通常使用柯达EDR2胶片和维达尔剂量测定PRO胶片数字化仪进行。然而,由于许多医院正朝着无胶片环境发展,可能无法使用胶片冲洗机。因此,我们研究了一种新推出的用于IMRT剂量测定的Gafchromic EBT胶片。将平面IMRT剂量分布同时传递到EBT和EDR2胶片上,并用维达尔VXR - 16以及爱普生Expression 1680平板扫描仪进行扫描。然后将测量的剂量分布与使用Pinnacle治疗计划系统计算的剂量分布进行比较。IMRT治疗包括用7 - 9束6兆伏的射线束治疗前列腺、头颈部以及其他一些部位。使用FilmQATM(3cognition有限责任公司)软件对胶片进行分析。测量剂量分布与计算剂量分布之间的比较报告为剂量差异(DD)(±5%范围内的像素)、符合距离(DTA)(3毫米)以及伽马值(γ)(剂量 = ±3%,距离 = 2毫米)。使用维达尔扫描仪搭配EDR2胶片是一种成熟的技术,计算剂量分布与测量剂量分布在所有指标(DD、DTA和伽马)上的一致性均优于90%。然而,对于在对数模式下用维达尔扫描仪进行的EBT胶片扫描,与计算结果的一致性变差,降至较低的80%。在线性模式下获得的EBT维达尔扫描结果显示一致性有所改善,达到较高的80%范围,但在整个扫描过程中仍观察到伪影。这些伪影在所有EBT扫描中都非常明显,可归因于胶片通过扫描仪的传输方式。在爱普生扫描仪中,两种胶片都被牢固固定,光源在胶片上扫描。结果发现,爱普生扫描仪对两种类型的胶片表现同样良好,在所有指标上的一致性均优于90%。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验