Abstract

Continuous flow blood pumps, such as axial flow and centrifugal pumps, have been gaining interest as circulatory devices for total artificial hearts (TAHs) and a biventricular assist device (BVAD) because of their smaller size and simpler structure compared to pulsatile pumps. However, continuous flow pumps are more prone to suction of the left ventricle than pulsatile pumps are. Sudden increases in flow rate to meet changes in physiological demand, especially in the left pump, often cause ventricle suction. In this study, a control algorithm to prevent suction from occurring in the left ventricle by controlling the rotational speed of the right pump, instead of reducing the cardiac output of the left pump, was developed and investigated. The method was tested in acute animal experiments with calves. The results indicate that this proposed method is capable of preventing suction and could simultaneously maintain circulatory control. A key advantage of this control system is that flow rates can be maximized while avoiding ventricle suction conditions particularly when the circulatory system is unstable such as in a the first few days after operation.