Medina Luis A, Goins Beth, Rodríguez-Villafuerte Mercedes, Bao Ande, Martínez-Davalos Arnulfo, Awasthi Vibhudutta, Galván Olga O, Santoyo Cristina, Phillips William T, Brandan María-Ester
Instituto de Fisica, UNAM, A.P. 20-364, Mexico 01000 D.F., Mexico.
Eur J Nucl Med Mol Imaging. 2007 Jul;34(7):1039-49. doi: 10.1007/s00259-006-0297-x. Epub 2007 Feb 8.
A procedure for the measurement of spatial dose rate distribution of beta particles emitted by 186Re-liposomes in tumoral tissue, using HS GafChromic films, is presented.
HNSCC xenografts were intratumorally injected with 3.7 or 11.1 MBq of 186Re-liposomes, and planar gamma camera images were acquired to determine the liposome retention in the tumor. After imaging, rats were sacrificed and tumors were excised and processed in slices; HS film sections were placed between slices and the tumor lobe was reassembled. Tumors and films were kept in the dark at 4 degrees C for 18 h. After irradiation, films were removed and response was read using a transmission scanner. Films were analyzed to determine two-dimensional spatial dose rate distributions and cumulative dose volume histograms. Dose rate distributions were quantified using a 60Co calibration curve, the 186Re physical half-life, and a perturbation factor that takes into account the effect of the film protective layer.
Dose rate distributions are highly heterogeneous with maximal dose rates about 0.4 Gy h(-1) in tumors injected with 3.7 MBq and 1.3 Gy h(-1) in tumors injected with 11.1 MBq. Dose volume histograms showed dose distributed in more than 95% and 80% of the tumor when injected with the lower and the higher activity, respectively.
The described procedures and techniques have shown the potential and utility of HS GafChromic film for determination of dose rate distributions in solid tumors injected intratumorally with 186Re-liposomes. The film's structure and the liposomes' biodistribution must be taken into account to obtain quantitative dose measurements.
介绍一种使用HS GafChromic胶片测量186Re脂质体在肿瘤组织中发射的β粒子空间剂量率分布的方法。
将3.7或11.1 MBq的186Re脂质体瘤内注射到头颈部鳞状细胞癌异种移植瘤中,采集平面γ相机图像以确定脂质体在肿瘤中的滞留情况。成像后,处死大鼠,切除肿瘤并切成薄片;将HS胶片切片置于薄片之间,重新组装肿瘤叶。肿瘤和胶片在4℃黑暗中保存18小时。照射后,取出胶片,使用透射扫描仪读取响应。分析胶片以确定二维空间剂量率分布和累积剂量体积直方图。使用60Co校准曲线、186Re物理半衰期和考虑胶片保护层影响的扰动因子对剂量率分布进行量化。
剂量率分布高度不均匀,注射3.7 MBq的肿瘤中最大剂量率约为0.4 Gy h(-1),注射11.1 MBq的肿瘤中最大剂量率约为1.3 Gy h(-1)。剂量体积直方图显示,注射较低和较高活度时,分别有超过95%和80%的肿瘤有剂量分布。
所描述的程序和技术已显示出HS GafChromic胶片在确定瘤内注射186Re脂质体的实体瘤剂量率分布方面的潜力和实用性。为获得定量剂量测量结果,必须考虑胶片结构和脂质体的生物分布。