一种测量呼吸系统静态压力-容积曲线的自动化方法及其在健康肺和油酸注入致肺损伤后的应用。

An automated method for measuring static pressure-volume curves of the respiratory system and its application in healthy lungs and after lung damage by oleic acid infusion.

作者信息

Zhao Y, Rees S E, Kjaergaard S, Smith B W, Larsson A, Andreassen S

机构信息

Center for Model-Based Medical Decision Support, Institute for Health Science and Technology, Aalborg University, Niels Jernes Vej 14, 4-311, Aalborg East, Denmark.

出版信息

Physiol Meas. 2007 Mar;28(3):235-47. doi: 10.1088/0967-3334/28/3/001. Epub 2007 Jan 23.

Abstract

Elastic pressure/volume (PV) curves of the respiratory system have attracted increasing interest, because they may be helpful to optimize ventilator settings in patients undergoing mechanical ventilation. Clinically applicable methods need to be fast, use routinely available equipment, draw the inspiratory and expiratory PV curve limbs, separate the resistive and viscoelastic properties of the respiratory system from the elastic properties, and provide reproducible measurements. This paper presents a computer-controlled method for rapid measurements of static PV curves using a long inflation-deflation with pauses, and its evaluation in six pigs before and after lung damage caused by oleic acid. The method is fast, i.e. 20.5 +/- 1.9 s (mean +/- SD) in healthy lungs and 17.7 +/- 4.1 s in diseased lungs, this including inspiratory and expiratory pauses of 1.1 s duration. In addition the only equipment used was a clinical ventilator and a PC. For healthy and damaged lungs expiratory PV curve limbs were very reproducible and were at higher volume than the inspiratory limbs, indicating hysteresis. For damaged lungs inspiratory PV limbs were reproducible. For healthy lungs the inspiratory limbs were reproducible but only after the first inflation-deflation. It is possible that during the first inflation alveoli are recruited which are not derecruited on deflation, shifting the inspiratory limb of the PV curve. The paused long inflation-deflation technique provides a quick, automated measurement of static PV curves on both inspiratory and expiratory limbs using routinely available equipment in the intensive care unit.

摘要

呼吸系统的弹性压力/容积(PV)曲线已引起越来越多的关注,因为它们可能有助于优化机械通气患者的呼吸机设置。临床适用的方法需要快速,使用常规可用设备,绘制吸气和呼气PV曲线分支,将呼吸系统的阻力和粘弹性特性与弹性特性分开,并提供可重复的测量结果。本文介绍了一种计算机控制的方法,用于通过长时间充气-放气并暂停来快速测量静态PV曲线,并在油酸致肺损伤前后的六头猪身上进行了评估。该方法速度快,即在健康肺中为20.5±1.9秒(平均值±标准差),在患病肺中为17.7±4.1秒,这包括持续1.1秒的吸气和呼气暂停。此外,唯一使用的设备是一台临床呼吸机和一台个人电脑。对于健康和受损的肺,呼气PV曲线分支非常可重复,并且容积高于吸气分支,表明存在滞后现象。对于受损肺,吸气PV分支是可重复的。对于健康肺,吸气分支是可重复的,但仅在第一次充气-放气之后。有可能在第一次充气期间募集了一些在放气时不会再塌陷的肺泡,从而使PV曲线的吸气分支发生偏移。这种暂停的长时间充气-放气技术可使用重症监护病房中常规可用的设备,对吸气和呼气分支的静态PV曲线进行快速、自动的测量。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验