Melnyk Roman, DiBianca Frank A
Department of Biomedical Engineering and Imaging, College of Medicine, University of Tennessee Health Science Center, Memphis, Tennessee 38163, USA.
Med Phys. 2007 Mar;34(3):1062-75. doi: 10.1118/1.2436977.
The detector presampling modulation transfer function (MTF) of a 576-channel variable resolution x-ray (VRX) computed tomography (CT) scanner was evaluated in this study. The scanner employs a VRX detector, which provides increased spatial resolution by matching the scanner's field of view (FOV) to the size of an object being imaged. Because spatial resolution is the parameter the scanner promises to improve, the evaluation of this resolution is important. The scanner's pre-reconstruction spatial resolution, represented by the detector presampling MTF, was evaluated using both modeling (Monte Carlo simulation) and measurement (the moving slit method). The theoretical results show the increase in the cutoff frequency of the detector presampling MTF from 1.39 to 43.38 cycles/mm as the FOV of the VRX CT scanner decreases from 32 to 1 cm. The experimental results are in reasonable agreement with the theoretical data. Some discrepancies between the measured and the modeled detector presampling MTFs can be explained by the limitations of the model. At small FOVs (1-8 cm), the MTF measurements were limited by the size of the focal spot. The obtained results are important for further development of the VRX CT scanner.
本研究评估了一台576通道可变分辨率X射线(VRX)计算机断层扫描(CT)扫描仪的探测器预采样调制传递函数(MTF)。该扫描仪采用了VRX探测器,通过将扫描仪的视野(FOV)与被成像物体的大小相匹配来提高空间分辨率。由于空间分辨率是该扫描仪承诺要提高的参数,因此对这种分辨率的评估很重要。使用建模(蒙特卡罗模拟)和测量(移动狭缝法)两种方法对以探测器预采样MTF表示的扫描仪重建前空间分辨率进行了评估。理论结果表明,随着VRX CT扫描仪的视野从32厘米减小到1厘米,探测器预采样MTF的截止频率从1.39周期/毫米增加到43.38周期/毫米。实验结果与理论数据合理吻合。测量的和建模的探测器预采样MTF之间的一些差异可以用模型的局限性来解释。在小视野(1 - 8厘米)下,MTF测量受焦点尺寸的限制。所获得的结果对VRX CT扫描仪的进一步发展具有重要意义。